F 406
Safety relay 1
1: Positive safety / 0: Negative safety
F 410
Activation mode for relay 2
0: None
1: Rising edge
2: Falling edge
3: Monitoring
F 411
Channel selection relay 2
Channel 1
Channel 2
F 412
Threshold 1 relay 2 setting
According to the connected probe
F 413
Threshold 2 or hysteresis relay 2 setting
According to the connected probe
F 414
Delay 1 relay 2 setting
From 0 to 600 seconds
F 415
Delay 2 relay 2 setting
From 0 to 600 seconds
F 416
Safety relay 2
1: Positive safety
0: Negative safety
10.5. F 500
Code
Transmitter
Description
Possibilities
F 500
CP210-R only
Integration of the measurement (pressure) From 0 to 9
F 501
CTV210-R only
Integration of the measurement (air
velocity)
From 0 to 9
F 510
CP210-R with
solenoid valve only
Time-delay between two auto-calibrations From 0 to 60 minutes
F 521
All
Offset main parameter
According to the measurement
range of the transmitter
F 531
All
Offset secondary parameter
According to the measurement
range of the transmitter
10.6. F 600
Code
Transmitter
Description
Possibilities
F 600
CP210-R
Temperature unit
0: °C
1: °F
F 602
CP210-R
Temperature compensation
F 605
COT212-R, CP210-R and
CTV210-R
Altitude compensation
F 610
CP210-R
Means of measurement
0: Pitot L
1: Pitot S
2: Debimo blade
3: Factor
F 611
CP210-R
Differential pressure
coefficient
F 612
CP210-R and CTV210-R
Factor of air velocity
correction
FUNCTIONS RECAP