0/3:
relay is deactivated
•
1/3:
rising edge mode
•
2/3:
falling edge mode
•
3/3:
monitoring mode
➢
Press
OK
.
“F 400”
blinks.
➢
Press
Up
key to go to the
“F 401”
folder of relay 1 (
“F 411”
for relay 2) then press
OK
.
➢
With
Up
and
Down
keys, select the channel where the relay will be activated.
➢
Press
OK
.
“F 401”
blinks.
➢
Press
Up
key to go to the
“F 402”
folder (
“F 412”
for relay 2) then press
OK
.
This folder corresponds to the
threshold 1
settings.
➢
Set the
threshold 1
with the
Up
and
Down
keys.
➢
Press
OK
when the last digit has been set.
“F 402”
blinks.
➢
Press
Up
key to go to the
“F 403”
folder (
“F 413”
for relay 2) then press
OK
.
This folder concerns:
•
For a rising or falling edge:
hysteresis setting
•
For a monitoring:
threshold 2 setting
➢
Set the hysteresis or threshold 2 with the
Up
and
Down
keys.
➢
Press
OK
when the last digit has been set.
“F 403”
blinks.
➢
Press
Up
key to go to the
“F 404”
folder (
“F 414”
for relay 2) then press
OK
.
This folder corresponds to the
delay 1
setting.
➢
Set the delay 1 with the
Up
and
Down
keys.
➢
Press
OK
when the last digit has been set.
“F 404”
blinks.
➢
Press the
Up
key to go to the
“F 405”
folder (
“F 415”
for relay 2) then press
OK
.
This folder corresponds to the
delay 2
setting.
➢
Set the delay 2 with
Up
and
Down
keys.
➢
Press
OK
when the last digit has been set.
“F 405”
blinks.
➢
Press
Up
key to go to the
“F 406”
folder (
“F 416”
for relay 2) then press
OK
.
This folder corresponds to the relay safety. The relay outputs are by default, in
negative security
: the relay is
energized when a set point is reached. With the keypad, you can swap the relays in
positive security
: then, the
relay is de-energised when a set point is reached or during a power outage.
➢
Activate or deactivate the safety relay with
Up
and
Down
keys.
•
1:
positive safety
•
0:
negative safety
➢
Press
OK
to confirm.
F 400: MANAGE THE RELAYS
17