background image

11.   Zero Range (4% to full capacity)

12.   Tare Range (100% to full capacity)

13.   Error Message indication

              EEEEE   Ovrload

       ErrL     Power on zero count too low

       ErrH     Power on zero count too high

       ErrE     Eeprom Error

    

      Low battery indication

14.

 

Three calibration weight units : kg , lb,oz.

15.   Open calibration capacity
16.   Power on zero-setting range: + 10%
17.   Zero range: + 4% of full capacity
18.   Three modes : Normal mode / Setting mode /Internal Auto Calibration mode
19.  

 

RS232 function

When to calibrate

Calibration  may  be  required  when  it  is  initially  installed,  if  the  scale/balance  is
moved to a substantial distance.
This is necessary because the weight of a mass in one location is not necessarily the
same  in  another  location.  Also,  with  time  and  use,  mechanical  deviations  may
occur.
Calibration Weight
1. Turn the power off. Press [UNITS] and [ZERO] first, then turn the power on.,
while  keeping  pressing  [UNITS]  and  [ZERO],  the  display  will  show  the  “CAL
ukg”.
2.  Press  [ZERO]  key,  the  display  will  show  AD  value,  press  [ZERO]  key,  the
display will show “SAVE”, and then show calibration value, place the calibration
value,  press  [ZERO]  key,  the  display  will  show  “SAVE”,  and  then  return  to  AD
value.
3. Now calibration weight is complete.

Summary of Contents for KPS 60

Page 1: ...e value 2 UNITS A Select weigh units kg lb or oz 3 TARE A When there s load on the scale press this key to return to zero B Put a container on the scale press this key to tare the weight and the tarred value is stored 4 HOLD A Press this key to store weight value 5 POWER SWITCH when the scale is used or not used please turn the power on or the power off ...

Page 2: ...ion Information Format show STX Start of Transmission 02H Status Status bit Example 2AH Stable weight 7EH Unstable Weight 2BH Positive Weight 2Dh Negat Weight DDDDDDD Value of Weight 7 ASCII Code UUUUU 4 ASCII Code Weight Units kg or LB ETX Stop of Transmission 03H SET UNIT CAP 3 DIVS 0 5 AOFF TARE UNIT CAL t b r GrAn 4800 9600 ON OFF Unit Unit Unit Unit Unit Unit Unit Unit Unit Unit TARE UNIT FCS...

Page 3: ...ng the local gravity 2 press TARE key again rightward flash digit 3 press HOLD key increase digit input the gravity value press UNITS the display enter to another local acceleration to gravity mode 3 When the display show U9 00 for flash digit repeat the steps 1 3 press UNITS the display will return the normal weighing mode Features 1 Compact Dimensions 355 405 60 Base 553mm x 470mm x 130mm 2 Mate...

Page 4: ...e required when it is initially installed if the scale balance is moved to a substantial distance This is necessary because the weight of a mass in one location is not necessarily the same in another location Also with time and use mechanical deviations may occur Calibration Weight 1 Turn the power off Press UNITS and ZERO first then turn the power on while keeping pressing UNITS and ZERO the disp...