KPC 3k and 15k
MAN08152022
03
Press MODE
“Ut ABC”
A =zero range when power on, 1~9 means 10%~90% FS
to zero.
B= zero tracking range, 1~9 means (1~9) x0.25d for
tracking range.
C=zero display according to the division setting, C=1
means when the weight<1d, display will show zero.
Press “→T←” to select position and press “→0←” key to
change data. *Press “MODE” to confirm and enter into
next step.
06
Press MODE
“bUF SAB”
Weighing speed response
S: 0=common; 1= speed up (AD internal calculation
speed.)
A: stable display parameter setting:
Setting range:0
、
1
、
2
、
3, response speed: 0>1>2>3.
B: weighing stable status display:
0=fast; 1=stable display; 2=slow; 3=after stable display.
Press “→T←” to select position and press “→0←” key to
change data. *Press “MODE” to confirm and enter
into next step.
07
Press MODE
“- HoLd -”
Weighing stable locking setting.
Press “→T←” enter into “FLTO X” weighing stable lock
setting.
X=0: unlocking;
X=1: locking after common stable;
X=2: press “→0←” to lock and unlock;
X=3: weight value of buzz locking, press “→0←” to
unlock.
Press “→0←” key to change data.
*Press “MODE” to confirm and enter into next step.
Press MODE
“FLT1 X”
Locking time setting (only FLTO=1 effective).
X=0 means after stable locking.
X=1~9 means after weighing stable 1~9 seconds locking.
Press “→0←” key to change data.
*Press “MODE” to confirm and enter into next step.
Press MODE
“FLT2 X”
Unlocking time setting (only FLTO=1 effective).
X=0 means when weight <20d, unlocking after stable.
X=1~9 means after weighing stable 1~9 seconds locking.
Press “→0←” key to change data.
*Press “MODE” to confirm and enter into next step.
08
Press MODE
“- InTE -
Weighing round number setting
Press →T←
“n1=00”
N1=00~99 (round number range) can be selected.
Press “→T←” to select position and press “→0←” key to
change data.
*Press “MODE” to confirm and enter into next step.
Press MODE
“n2=0”
Valid time setting.
Over this valid time, real weight will display.
*Press “MODE” to confirm and enter into next step.