4-2 Using Terminals and Connectors
TOS9200/9201
4.1
REMOTE Terminal
The REMOTE terminal is the 5-pin DIN connector on the front panel.
This connector is used exclusively to connect the optional remote-control box
RC01-TOS/RC02-TOS and high-voltage test probe HP01A-TOS/HP02A-TOS.
When this optional device is connected, the START switch on the front panel
becomes invalid.
Fig.4-1 REMOTE terminal
•
Use the optional high-voltage test probe HP01A-TOS/HP02A-TOS at AC 4
kV or DC 5 kV or less.
•
Keep the signal line at least 500 mm from the high-voltage test leadwire
and the DUT. Do not short-circuit the test voltage to the signal line. Other-
wise, all internal circuits may be destroyed.
Control procedure
1.
Turn off the POWER switch.
2.
Using the exclusive connection cable (5-pin DIN cable), connect the
optional device to the REMOTE connector on the front panel.
3.
Turn on the POWER switch.
A test can be started on the optional device. In such a case, the START switch
on the panel becomes invalid. To stop the test, either press the STOP switch on
the panel or input the STOP signal on the optional device. For details, see the
operation manual for the optional device.
4.
To return to the panel for control,
fi
rst turn off the POWER switch on the
tester.
5.
On the front panel, remove the exclusive connection cable (5-pin DIN
cable) from the REMOTE connector.
6.
Turn on the POWER switch.
The START switch on the panel becomes valid.
REMOTE terminal
WARNING
CAUTION
Summary of Contents for TOS9200
Page 22: ...P 10 Preface TOS9200 9201...
Page 116: ...3 78 Basic Operations TOS9200 9201...
Page 130: ...4 14 Using Terminals and Connectors TOS9200 9201...
Page 148: ...7 4 Maintenance TOS9200 9201...
Page 162: ...8 14 Specifications TOS9200 9201...
Page 170: ...I 4 Index TOS9200 9201...
Page 171: ......