34 Panel Operations
TOS8030
5.3
Procedure for Test
5.3.1
Test Parameters
The following parameters must be set.
Set these values according to the standards of the withstanding voltage test to be
performed.
Current detection reference value
If the measured leakage current exceeds the value set here, a FAIL determination is
returned. Valid current values range from 0.1 mA to 11 mA. Setting the value to 0
mA or 11 mA or higher will cause the tester to enter a PROTECTION state (P07 or
P06).
Fig. 5-6
Setting a Current Detection Reference Limit
• If the operator sets a current value exceeding 5 mA, testing is subject to output
time limitations. For more information, see “Footnote *1” in “9.1 Basic Specifica-
tions” (page 46).
Parameter
Range
Current detection reference value
0.1 mA to 11 mA
Test time
TIMER OFF, 0.5 s to 99 s
Test voltage
0.05 kV to 3.00 kV
Valid current values:
×
0.1 mA range: 0.1 mA to 9.9 mA
×
1 mA range: 1 mA to 11 mA
Select the value range.
Turn up or down to set reference current values (mA).
If the
×
0.1 mA range is selected,
“05” signifies a setting of 0.5 mA.
NOTE