background image

PLZ-5W

Connecting to the Device Under 

Test (Cont'd)

Connecting to the load input terminals on the 

front panel

The load input terminals on the front panel enable you to 

easily connect the device under test (DUT) to the PLZ-5W. The 

specifications of the PLZ-5W are for the load input terminals on 

the rear panel and the load input terminals on the front panel 

may not meet the specifications.

WARNING

• Risk of electric shock. Attach insulation caps to the 

crimping terminals.

• The rated current value of the load input terminals on 

the front panel of the PLZ1205W is 80 A. When a current 

of 80 A or more flows to the load input terminals on the 

front panel, an alarm occurs and load off.

 

Turn the POWER switch off ( ).

 

Make sure that the output of the DUT is off.

 

Remove the protection plate of the load input 

terminals on the front panel.

Keep the removed protection plate and screws in a safe 

place. By attaching the protection plate when not using the 

load input terminals on the front panel, the PLZ-5W can be 

used in a safe and compact manner.

 

Attach the cover for the load input terminals on 

the front panel.

 

Insert the load cables from the bottom and tight-

en the accessory knobs to fix the cables in place.

knob

knob

Black

(-)

Black

(-)

Red

(+)

Red

(+)

 

Connect the load cables to the output terminals 

of the DUT.

Connect the positive (+) polarity of the load input terminal on 

the front panel to the positive (+) polarity of the DUT, and the 

negative (

) polarity of the load input terminal on the front 

panel to the negative (

) polarity of the DUT.

Connecting the Sensing Cables

To perform remote sensing, connect the sensing cables.

A voltage measurement point can be changed from a load input 

terminal to an arbitrary sensing point by executing remote sens-

ing. By setting sensing points to a DUT end, influences such as 

voltage drops caused by the resistance of the load cables can 

be reduced and the load current can be stabilized.

CAUTION

• Risk of damage to the internal circuitry. Never wire the sensing 

terminals while the POWER switch is turned on.

• If the remote sensing cables come loose while remote sensing 

is in use, the PLZ-5W and DUT may be damaged. Be sure to 

connect the cables securely.

Unlike the load cables, there is no need to consider the cur-

rent capacity for the sensing cables. However, for adequate 

mechanical strength, use wires whose nominal cross-sectional 

area is 0.5 mm2 or greater.

 

Turn the POWER switch off ( ).

 

Remove the sensing terminal cover on the rear 

panel.

M3×6

M3×6

 

Connect the sensing cables to the PLZ-5W.

Use the screws to fix the 

wires in place.

We recommend stripping 

7 mm of the insulation 

from the wires.

7mm

+

 

Attach the sensing terminal cover.

 

Connect the sensing cables to the DUT.

Connect the positive (+S) polarity of the SENSING terminal 

to the positive (+) polarity of the DUT, and connect the neg-

ative (−S) polarity of the SENSING terminal to the negative 

(−) polarity of the DUT. Connect the cables as close to the 

EUT as possible.

DUT

PLZ-5W

+S

−S

Remote sensing terminal

+

+

Red

White

Twist

To use remote sensing, you need to enable the remote 

sensing feature. For details, see the User’s Manual on the 

included CD-ROM.

Summary of Contents for PLZ1205W

Page 1: ...ved from 1 5 V to 1 0 V offering support of low voltage input Arbitrary I V characteristics mode Arbitrary I V characteristics mode that allows current to be set arbitrarily in relation to voltage inp...

Page 2: ...e of how to control measuring instruments using a personal computer Quick reference This document briefly explains the control panel and the basic operation of it Safety information This document cont...

Page 3: ...ge Category II energy consuming equipment that is supplied from a fixed installation 1 Turn the POWER switch off 2 Check that the AC power line meets the nominal input rating of the PLZ 5W The PLZ 5W...

Page 4: ...nt flows and the voltage at the load input terminals may fall below the minimum operating voltage of the PLZ 5W Using the following table as a reference select wiring whose nominal cross sectional are...

Page 5: ...her There are two available positions Use the appropriate position for the load cables that you are using For cables up to 10 mm Put the top and bottom load input terminal covers together so that the...

Page 6: ...erminal on the front panel to the negative polarity of the DUT Connecting the Sensing Cables To perform remote sensing connect the sensing cables A voltage measurement point can be changed from a load...

Page 7: ...e PLZ 5W remotely by sending commands For details on remote control see the Communication Interface Manual on the included CD ROM The following environment is required to view the Communication Interf...

Page 8: ...demarks or trademarks of Microsoft Corporation in the United States and or other countries Other company names and product names used in this manual are generally trademarks or registered trademarks o...

Reviews: