background image

4 PLZ-5W

Checking Whether the Power Is  

On or Off

Turning the power on

POWER switch

 

Check that the power cord is connected correctly.

 

Check that nothing is connected to the DC INPUT 

(load input) terminals on the front and rear pan-

els.

 

Turn the POWER switch on ( ).

The display lights.

─ Note ─
If you notice strange sounds, unusual odors, fire, or smoke 

around or from inside the PLZ-5W, flip the POWER switch off, or 

remove the power cord plug from the outlet.

Turning the power off

 

Press the ( ) side of the POWER switch to turn 

the power off.

CAUTION

After you turn the POWER switch off, wait at least 5 seconds 

after the fan stops before you turn the POWER switch back 

on. Repeatedly turning the POWER switch on and off at short 

intervals will shorten the service life of the POWER switch and 

the internal input fuse.

Selecting the load cables

WARNING

• 

Risk of fire. Use load cables whose capacity is adequate 

for the PLZ-5W’s rated output current.

• Risk of electric shock. Use load cables with a voltage 

rating that meets or exceeds the isolation voltage 

(±500 V) of the PLZ-5W’s load input terminals.

CAUTION

Use load cables with a core diameter that is appropriate for the 

amount of current being used and with sturdy, flame-resistant 

insulation.

If the wiring that you use for the load has a high resistance, 

the voltage will drop significantly when current flows, and the 

voltage at the load input terminals may fall below the minimum 

operating voltage of the PLZ-5W. Using the following table as a 

reference, select wiring whose nominal cross-sectional area is 

as thick as possible.

Nominal 

Cross-Sectional 

Area [mm2]

AWG

(Reference 

Cross-Sectional 

Area) [mm2]

Allowable Current* 

[A](Ta = 30 °C)

14

6

(13.3)

88

22

4

(21.15)

115

30

2

(33.62)

139

38

1

(42.41)

162

50

1/0

(53.49)

190

60

2/0

(67.43)

217

80

3/0

(85.01)

257

100

4/0

(107.2)

298

125

-

  -

344

150

-

  -

395

200

-

  -

469

250

-

  -

556

325

-

  -

650

* Excerpt from Japanese laws related to electrical equipment.

Connecting to the load input terminals on the 

rear panel

WARNING

Risk of electric shock.

• Do not touch load input terminals when the output is 

turned on.

• The load input terminals on the front panel are con-

nected internally to the load input terminals on the rear 

panel. The voltage applied to the terminal on one side is 

transmitted directly to the terminal on the other side.

• Be sure to use the load input terminal cover.

CAUTION

To avoid damaging the product, observe the following precau-

tions.
• Do not connect the DUT to the load input terminals while the 

PLZ-5W’s load is turned on.

• Do not connect electronic loads to the load input terminals on 

the front panel and those on the rear panel at the same time.

• Do not invert the polarity when connecting. An overcurrent 

might flow when the load is turned on.

To avoid overheating, observe the following precaution.
• Attach crimping terminals to the cables, and use the supplied 

screws to connect the cables.

Connecting to the Device Under Test

Summary of Contents for PLZ1205W

Page 1: ...ved from 1 5 V to 1 0 V offering support of low voltage input Arbitrary I V characteristics mode Arbitrary I V characteristics mode that allows current to be set arbitrarily in relation to voltage inp...

Page 2: ...e of how to control measuring instruments using a personal computer Quick reference This document briefly explains the control panel and the basic operation of it Safety information This document cont...

Page 3: ...ge Category II energy consuming equipment that is supplied from a fixed installation 1 Turn the POWER switch off 2 Check that the AC power line meets the nominal input rating of the PLZ 5W The PLZ 5W...

Page 4: ...nt flows and the voltage at the load input terminals may fall below the minimum operating voltage of the PLZ 5W Using the following table as a reference select wiring whose nominal cross sectional are...

Page 5: ...her There are two available positions Use the appropriate position for the load cables that you are using For cables up to 10 mm Put the top and bottom load input terminal covers together so that the...

Page 6: ...erminal on the front panel to the negative polarity of the DUT Connecting the Sensing Cables To perform remote sensing connect the sensing cables A voltage measurement point can be changed from a load...

Page 7: ...e PLZ 5W remotely by sending commands For details on remote control see the Communication Interface Manual on the included CD ROM The following environment is required to view the Communication Interf...

Page 8: ...demarks or trademarks of Microsoft Corporation in the United States and or other countries Other company names and product names used in this manual are generally trademarks or registered trademarks o...

Reviews: