PLZ-4W
Basic Operation 5-5
When the DC power supply and the DUT are connected in series,
and turning on/off of the DUT. (PLZ164WA/PLZ664WA only)
The following describes for testing the DUT (in CC mode) when the DUT (an open-
close device such as a switch) is connected between the load input terminal of the
PLZ-4W and the output of the DC power supply.
Fig.5-1
Connecting the DC power supply and the DUT in series
As for the type of 0 V input model PLZ164WA/PLZ664WA, when the connection
between the DC power supply and the PLZ164WA/PLZ664WA is opened circuit, a
reverse voltage is applied to the load input terminal, and the alarm is occurred to
turn off the load by the reverse-connection detection (REV). For the REV, see sec-
tion 5.3, “Types of Protection Functions.”
■
To avoid the occurance of alarm by the reverse-connection
detection (REV)
To avoid the occurrence of alarm by the REV, set the operation mode to the
"CC+CV" mode, then it is necessary to set the voltage value of the CV at lower than
the voltage "V" speci
fi
ed in the Fig. 5-1 when the current
fl
ows. The alarm can not
be avoided in the CC mode. For the operation mode, see section 5.5, “Operation
Modes.”
1.
Turn off the DUT.
2.
Set the operation mode of the PLZ-4W to the "CC+CV" mode.
3.
Set the voltage of the CV for approximately 1 V lower than the output
voltage of the DC power supply.
For the setting of voltage, see “Operating procedure of CC+CV mode” in sec-
tion 5.6, “CC Mode.”
4.
Set the current of the CC complied to the condition of the DC power supply.
5.
Press the LOAD key to turn on the load.
6.
Turn on the output of the DC power supply.
7.
Turn on the DUT.
8.
Turn off the DUT when the test is
fi
nished.
Even when you follow the procedure speci
fi
ed above, the alarm may occurs, in such
case, it may be required to change the voltage setting of CV to prevent the alarm
occurance. For details, contact your Kikusui agent or distributor.
+
-
PLZ164WA
PLZ664WA
DC power
supply
DUT
Load input
terminal
v
Summary of Contents for PLZ-4W Series
Page 10: ...VIII Contents PLZ 4W ...
Page 20: ...1 10 General Information PLZ 4W ...
Page 38: ...2 18 Installation and Preparation PLZ 4W ...
Page 90: ...5 30 Basic Operation PLZ 4W ...
Page 192: ...9 10 Specifications PLZ 4W ...
Page 214: ...A 22 Appendix PLZ 4W ...