PLZ-4W
Basic Operation 5-11
4.
Select the voltage range.
Press the VRANGE (SHIFT+RANGE) key.
A voltage range pop-up menu appears.
Each time you press the VRANGE key, the voltage range switches
between 15 V and 150 V.
When the desired range is highlighted, stop pressing the key.
5.
Set the current value.
Check that the display is in the basic setting entry condition (the characters
“SET” is highlighted).
While viewing the display, turn the rotary knob.
First, use coarse adjustment to roughly set the value, and then switch to
fi
ne
adjustment to adjust the value precisely.
6.
Turn on the load.
Press the LOAD key.
The LOAD LED illuminates, and the current
fl
ows.
The display shows the measured values (voltage, current, and wattage)
at the load input terminal.
7.
Change the current setting.
Turn the rotary knob to change the current setting.
The current value cannot be set greater than the maximum value of the selected
range.
8.
Turn off the load.
Press the LOAD key.
The LOAD LED turns off, and the current is cut off.
■
Operating procedure of CC+CV mode
Fig. 5-4
Keys used in CC+CV mode
1.
Check the operation mode.
Check that CC is indicated on the display.
2.
Select the operation mode.
Press the +CV (SHIFT+MODE) key.
The mode switches to CC+CV.
Check that CC+CV is indicated on the display.
SWITCHING
ENTER
SHORT
EDIT
SLEW RATE
RUN / STOP
UVP
PAUSE
STORE
RECALL
A
B
C
LOAD
Th/TL
FREQ / DUTY
MENU
PRESET SEQ
I
LEVEL
VRANGE
RANGE
SW ON
+CV
MODE
LOAD
SET / VSET
LOCK
SHIFT
ABC
OPP / OCP
Summary of Contents for PLZ-4W Series
Page 10: ...VIII Contents PLZ 4W ...
Page 20: ...1 10 General Information PLZ 4W ...
Page 38: ...2 18 Installation and Preparation PLZ 4W ...
Page 90: ...5 30 Basic Operation PLZ 4W ...
Page 192: ...9 10 Specifications PLZ 4W ...
Page 214: ...A 22 Appendix PLZ 4W ...