Preparation | Preparing for Connecting a DUT
Taking measures against noise
When connecting cables that have the same heat resistance, separating the cables as much as possible
to increase heat radiation enables a greater amount of current to flow. However, wiring the + (positive) and
- (negative) output wires of the load cable side by side or bundling them together is more effective against
unwanted noise. The Kikusui-recommended currents shown in the table
are allowable currents that
have been reduced in consideration of the potential bundling of cables. Use these values as a guideline
when connecting cables.
Limitations of the remote sensing function
All wires have resistance. As the wire becomes longer or the current increases, a drop in voltage across
the wire becomes bigger. That means the voltage applied to the DUT terminal becomes small. This
product has a sensing function that compensates for this voltage drop up to approximately 1 V for a single
line. If the voltage drop exceeds this level, use cables that have a greater cross-sectional area.
Constructing cables
The optional set of output cables, TL13-PFX and TL14-PFX, are available
. The provided necessary
processing of the cables facilitates the connection, saving wiring time. We recommend using the optional
set of output cables to connect a DUT.
For information about options, contact your Kikusui agent or distributor.
Cables and tools that are necessary to make the connection
Output cable
(stranded wire)
Cable for running charge/discharge current between the PFX2731S and DUT
See the table related to electrical equipment
Connector element: DF22-2S-7.92C (28) made by Hirose Electric
Voltage sensing cable
Cable for sensing the electrode voltage of the DUT
AWG24 to AWG20 (0.20 mm
2
to 0.52 mm
2
), shielded cable
Connector element: FK-MCP1.5/5-STF-3.81 made by Phoenix Contact
“Constructing sensing cable”
Thermocouple
Thermocouple used for measuring DUT temperatures
Type T, Insulation material: Teflon, Element-wire diameter: 0.32 mm
2
Operating temperature range: -40 °C to +200 °C
See “Constructing temperature measurement cable”
Terminal (pin)
DUT connector terminal (pin): DF22A-1012SCA made by Hirose Electric
Flat-blade screwdriver
Core diameter:
Ø
3, blade tip width: 2.6 mm
Wire stripper
A wire stripper that matches the wires listed above
Manual crimping tool
HT103/DF22A-1012 made by Hirose Electric