6-6
PAN-A
Current system adjustment procedure
The current system includes the following three items. Since all the
items are related to one another, adjust them all in the following pro-
cedure.
• Output current offset
• Output current in full scale
• Output current display in full scale
■
Connection of equipment
1. Turn off the OUTPUT switch.
2. Turn off the POWER switch.
3. Check that the SENS switch is turned off.
If the SENS switch is turned on, it is impossible to adjust cor-
rectly.
4. Connect a shunt resistor and DVM as shown in Fig.6-4.
For the PAN16-50A user, see Fig.6-5.
5. Connect the - (neg.) terminal and the chassis ground termi-
nal (
⊥
) with the shorting bar.
• Use load cables of suf
fi
cient current capacity to cover
the rated current.
Fig.6-4
Connection for current system adjustment
WARNING
+
-
10A
Rear OUTPUT
terminal board
Front auxiliary output terminals
Crimped
terminal
Connect the shorting
bar to the - (neg.) terminal.
DVM
HI
LO
Shunt resistor