KJM6710
RS-232C Control 4-11
(Example) Set the symmetry mode to AUT.
SYM:MOD AUT
Response message
Returns the current setting in response to SYM:MOD? in charac-
ters (AUT, OFFS, or MAN).
SYMMETRY:OFFSET:LEVEL (SYM:OFFS:LEV)
Sets the symmetry offset or slice level. Or, queries the current set-
ting.
This message is valid only when the symmetry mode is OFFSET
or MANUAL. The speci
fi
ed value is applied to the symmetry off-
set or slice level depending on the symmetry mode.
Program message
• Syntax
Command message:SYMMETRY:OFFSET
:LEVEL_<value>
SYM:OFFS:LEV_<value>
Query message:
SYMMETRY:OFFSET:LEVEL?
SYM:OFFS:LEV?
• Program data
Data format:
Integer
Values:
–128 to +127
Resolution:
1
(Example) Set the symmetry offset or slice level to 32.
SYM:OFFs:LEV 32
Response message
Returns the current setting in response to SYM:OFFS:LEV? in
integers (–128 to +127).
SYMMETRY:SLICE:LEVEL (SYM:SLIC:LEV)
Queries the symmetry slice level.
The amplitude of the input signal (100 %) corresponds to a slice
level value of approximately 600. Thus, the returned value is
between –300 and +300.