HP01A-TOS/HP02A-TOS
5
Inspection Before Starting
Always inspect and confirm the following items before beginning an actual test:
* If any abnormality or failure is found in the “Inspection Before Starting,” immediately
stop using the test probe and contact your Kikusui distributor or agent for repairs.
Stop Use in the Event of Failures
If any of the following conditions arises, the test probe is defective. To minimize haz-
ards, immediately turn off the POWER switch of the Tester and stop using the test
probe.
• The H.V ON LED does not light.
• The on/off status of the H.V ON LED of the test probe is inconsistent with that of
the DANGER (TEST ON or H.V ON) lamp on the Tester.
• The H.V ON LED remains lit even when start switch 1 or 2 is turned off.
• The H.V ON LED remains lit even when the contact pin is retracted into the red
area.
Item
Description of Inspection
Remote control cable
Chack for broken sheathing and loose or cracked connectors.
High voltage cable and
LOW test lead
Check for broken sheathing and loose or cracked connectors.
Use a tester to confirm electrical continuity.
Grip and slide lever
Confirm that the slide lever moves smoothly.
Start switches 1 and 2
Confirm that they operate smoothly when pressed.
Contact pin
Check for any signs of damage.
Warning label
Check for illegibility or peeling.
WARNING
• Implement measures so that no one can use the test probe
until it has been repaired.
• To minimize potential hazards, always contact your Kiku-
sui distributor or agent for repairs.