HP01A-TOS/HP02A-TOS

3

To Supervisor in Charge of Operations

• Use of the test probe involves unavoidable electrical shock hazards. Take appro-

priate measures to ensure maximum safety in the test environment at the DUT, at 
the test bench, and in nearby areas.

• If the operator does not read the language used for this manual, translate the 

manual into an appropriate language.

• Ensure that the operator understands the information in this manual before start-

ing operations.

• Keep this manual near the product to allow for easy reference by the operator 

when necessary.

About Manuals

Intended readers

These manuals are intended for users of this product and their instructors. The man-
uals assume that the reader has knowledge about electric safety testing.

Trademarks

Company names and product names used in this manual are generally trademarks 
or registered trademarks of the respective companies.

Copyright

Reproduction and reprinting of this operation manual, whole or partially, without our 
permission is prohibited. Both unit specifications and manual contents are subject 
tochange without notice. 
© Copyright 2018 Kikusui Electronics Corporation

Summary of Contents for HP01A-TOS

Page 1: ...electric shock that may result in serious injury or death Never perform testing when leaning on your elbows at the test bench or in other areas Do not fasten the slide lever of the test probe such th...

Page 2: ...cting electric shock Modification may cause electric shock Never fasten the slide lever to the grip such that the contact pin is exposed Do not lock start switch 1 and or start switch 2 to the on stat...

Page 3: ...ep this manual near the product to allow for easy reference by the operator when necessary About Manuals Intended readers These manuals are intended for users of this product and their instructors The...

Page 4: ...sor in Charge of Operations 3 About Manuals 3 Inspection Before Starting 5 Stop Use in the Event of Failures 5 Component Names and Functions 6 Connecting to the Tester 7 Starting and Stopping Tests 8...

Page 5: ...V ON lamp on the Tester The H V ON LED remains lit even when start switch 1 or 2 is turned off The H V ON LED remains lit even when the contact pin is retracted into the red area Item Description of I...

Page 6: ...witch with a finger from your non dominant hand This hand should be free since the other hand holds the test probe 5 H V ON LED This red LED indicates that high voltage is being applied to the contact...

Page 7: ...r 4 Connect the high voltage cable to the high voltage terminal of the Tester Cable Lock The cable lock guard is not used Insert the tester side end of the test leadwire fixture into the terminal and...

Page 8: ...g test Disabled Press start switches 1 and2 Stopping test Press the STOP switch Release your finger from start switch 1 or 2 CAUTION Initiation of contact or removal from the DUT while high voltages a...

Page 9: ...t lit g 11Release your fingers from the slide lever to retract the contact pin into the red area h 12Release the finger of your non dominant hand from start switch 2 i 13Remove the test probe from the...

Page 10: ...ntly wipe the probe Periodic Inspection The remote control cable high voltage cable and LOW test leadwire are consum ables We recommend a inspection by one of our authorized service engineers once a y...

Page 11: ...d to the contact pin during testing It also includes areas where high volt age is electrostatically or electromagnetically induced by application of high voltages 2 Rated maximum voltage The rated max...

Page 12: ...ery effort has been made to ensure the accuracy of this manual However if you have any questions or find any errors or omissions please contact your Kikusui agent or distributor After you have finishe...

Reviews: