Measurer : User Name

Test date : 2019.01.23.09:00

Bench mark value 1 (Pass) 200
Bench mark value 2 (Caution) 400

Pass

kitchen knife

Result

Select the measurement 

point and start 

measurement with the 

application.

Summary of Contents for Lumitester Smart

Page 1: ... immediately start hygiene management Measure Start Recommended list of measurement targets in the kitchen choice Choice Thank you very much for purchasing our product This Instruction Manual must be read carefully to operate this product safely and properly This Instruction Manual should be kept for future reference Measurer User Name Test date 2019 01 23 09 00 Bench mark value 1 Pass 200 Bench m...

Page 2: ...ing allows data sharing by remote sites The measurement result is easy to see Data are stored in the cloud Measurement workflow Package contents The trend history of the test point is displayed to check the improvement effect at a glance View Save 1 Lumitester Smart 1 Strap 2 size AA alkaline batteries for operation check 3 cleaning brushes 1 USB cable 1 Quick manual Pass Start kitchen knife Swab ...

Page 3: ...ent procedure 19 4 3 Measurement data check 23 5 Product Setup 24 5 1 BLE Pairing Mode Connection to 2nd or later smartphone tablet 25 5 2 Temperature Compensation Temperature compensation setting 26 5 3 Self Check Self diagnosis 27 5 4 Switch Sensitivity Switch sensitivity adjustment 29 5 5 Contrast Display contrast adjustment 29 5 6 Information Product information 30 5 7 Data Clear Measurement d...

Page 4: ...ver warning Indicates communication with smartphone tablet and or PC Indicates connection with smartphone tablet and or PC Indicates that the Bluetooth ON OFF switch of the product is ON Displays the remaining battery level The measured value is displayed up to 6 digits with right justification Shows the measured data number with 4 digits Indicates that the temperature compensation setting is ON I...

Page 5: ...es It could cause re injury or failure Remove the batteries when the product is to be stored for an extended period of time Failure to do so may cause liquid leak and burst Do not place the product where it gets hot near the fire near the heater or under the scorching sun Fire burn or failure may occur Do not push nor rub the display and operation panel with hard or sharp objects Injury or failure...

Page 6: ...er measurement Failure to do so may lead to liquid leak Do not use nor store the product in locations subject to large variations in temperature Do not use nor store the product in locations directly exposed to wind from air conditioning equipment Allow the product to stand for 30 minutes or longer at room temperature before use when the product is moved from a hot or cold location Otherwise measu...

Page 7: ...Bluetooth cannot be used to connect to a PC Please use the provided USB cable Caution Place the batteries Replacement of batteries 5 seconds after remaining battery amount is displayed with low battery the power is automatically turned off Replace the batteries After removing the batteries the clock count continues for 60 seconds Please replace the batteries within 60 seconds If Bluetooth is not u...

Page 8: ...nection to smartphone tablet pairing Use Environment Supported OS and model as of Sep 2018 Smartphone or tablet with Android 5 x or later or iOS 10 x or later Turn ON the Bluetooth of the smartphone or tablet Run the Application and follow the instructions on the screen to set up On the device connection screen shown turn on the power of the product If you use the Application for the first time pl...

Page 9: ...Application refer to Help To connect the 2nd and subsequent smartphones and or tablets refer to 5 Product Setup 5 1 BLE Pairing Mode Connection to 2nd or later smartphone tablet P25 Caution Connection between smartphone and or tablet and the product for the 2nd time onward On the Application s Home Screen open the menu Select the icon to display the connection screen with the product Press and hol...

Page 10: ...t Supported OS and model as of Sep 2018 PC or tablet with Windows 7 32 bit 64 bit or Windows 10 32 bit 64 bit installed This Software allows data transfer data from the product to your PC and set up various measurement conditions from your PC to the product Do not use this product for any other application Insert the batteries to the product before connecting it with a PC Without the batteries the...

Page 11: ...r the product Insert the batteries before connecting the product with a PC The display panel shows USB and key operation is disabled while the product is connected to a PC When you pull out the USB cable the display USB is cleared and the power is turned off Close the USB cover securely when the USB cable is not connected Caution USB Re connection to PC once connected 0004 USB Double click Lumites...

Page 12: ...ayed and it is ready for measurement after countdown Caution 4 Operation Method 0004 Lumitester 4 Remove the swab stick of the reagent Soak the cotton swab in tap water washing water and swab the test target Put the swab stick back into the main tube And then put the tip of the tube on your hand and push the swab stick downwards to break the capsule in the tube Shake the whole reagent tube repeate...

Page 13: ... may cause malfunction or interfere with accurate measurement When the reagent warning blinks and the buzzer sounds after pressing START button no reagent has been placed in the measurement chamber In order to cancel the measurement press START button again When the reagent warning blinks and the buzzer sounds after measurement the reagent has been placed in the measurement chamber Open the measur...

Page 14: ...ct a setup item and press START Pressing the Power button terminates the setup and the screen returns to the measurement screen Setup Menu Normally memory data is not erased while the power is off However when the measurement is continued with low battery level or when the batteries are removed while the power is on the memory data may be erased We are not responsible for damage caused when memory...

Page 15: ...ture of the product The temperature compensation is OFF by default The range of the temperature compensation is 10 to 40 C 5 2 Temperature Compensation Temperature compensation setting Press and hold the Power button for 5 seconds or longer and display the Setup screen Then press Up Down button to select Temperature Compensation and press START Select availability of temperature compensation using...

Page 16: ...ning the measurement chamber cover starts countdown and the interior of the measurement chamber emits light Clean the measurement chamber Maintenance of measurement chamber 1 Thoroughly wipe the locations that emit light in the measurement chamber with the provided cleaning brush soaked with ethanol 2 Close the measurement chamber cover To make sure that the measurement chamber is clean start the ...

Page 17: ...t You can adjust sensitivity of START and Up Down buttons 5 5 Contrast Display contrast adjustment You can adjust the contrast of the display panel Press and hold the Power button for 5 seconds or longer and display the Setup screen Then press Up Down button to select Information Press and hold the Power button for 5 seconds or longer and display the Setup screen Then press Up Down button to selec...

Page 18: ...hamber approximately once every six months Perform self check as needed For self check refer to 5 3 Self Check P27 Turn the power off Open the measurement chamber cover Thoroughly wipe the bottom and the side of the measurement chamber with the provided cleaning brush soaked with ethanol Close the measurement chamber cover 6 Maintenance 6 1 Maintenance of the main body If the product becomes conta...

Page 19: ...n remove the batteries When abnormality is observed such as malfunction burning smell fuming etc there is a danger of fire and burst Make sure that fumes are extinguished and contact the dealer or us Never repair the product by yourself as this is very dangerous Warning Do not mistake the polarities of batteries Do not mix batteries of different manufacturers or models or new and used batteries Do...

Page 20: ...asures are described below When the error code is still displayed after countermeasures are taken check the model and serial No affixed inside the battery cover and contact the dealer or us E011 to 019 Memory error Turn the power off immediately and remove the reagent Wipe off water with the cleaning brush and dry the product with the operation panel turned upward and the measurement chamber cover...

Page 21: ...type 0 to 999999 RLUs 10 to 40 C 10 seconds If the Temperature compensation is ON the measurement time at 10 to 13 C is 20 seconds Organic LED Built in 10 minutes Built in date and time Valid only when the Application is connected automatic synchronization RLU USB Bluetooth LE Ver 4 1 2000 5 to 40 C 20 to 85 Rh free from condensation 10 to 50 C 20 to 90 Rh free from condensation IEC IP X0 protecti...

Page 22: ... or using the product contrary to what is described in the instruction manual being careless when using modifying or any other remodeling of this product or being affected by a natural disaster beyond human control The scope of indemnification for failure of this product is limited to indemnification covered by this warranty section and indemnification of any other loss or damage including indirec...

Page 23: ...man Corporation Windows is a registered trademark of Microsoft Corporation U S A Bluetooth is a registered trademark owned by Bluetooth SIG Inc Kikkoman Corporation uses them based on the licenses Issued by http biochemifa kikkoman co jp e 2019 01 59 2756 0 ...