30
Keysight E1463A 32-Channel Form C Switch Service Guide
Verification Tests
Performance Verification
590ELSE
600OUTPUT @Sw;"OPEN (@"&Cc$&VAL$(I)&")"
610END IF
620END IF
630OUTPUT @Sw;"*OPC?"
640ENTER @Sw;Opc
650IF Opc<>1 THEN GOTO 630
660OUTPUT @Dmm;"TRIG SGL"
670ENTER @Dmm;Result(J,I)
(Continued on Next Page)
680 IF J<1 THEN
690 IF I <10 THEN
700 OUTPUT @Sw;"OPEN (@"&Cc$&Ph$&VAL$(I)&")"
710 ELSE
720 OUTPUT @Sw;"OPEN (@"&Cc$&VAL$(I)&")"
730 END IF
740 ELSE
750 IF I <10 THEN
760 OUTPUT @Sw;"CLOS (@"&Cc$&Ph$&VAL$(I)&")"
770 ELSE
780 OUTPUT @Sw;"CLOS (@"&Cc$&VAL$(I)&")"
790 END IF
800 END IF
810 OUTPUT @Sw;"*OPC?"
820 ENTER @Sw;Opc
830 IF Opc<>1 THEN GOTO 810
840 IF Result(J,I)<0 THEN Result(J,I)=0
850 IF Result(J,I)>2.0 THEN
860 PRINT "Resistance for Channel ";I;" "&Path$(J)&" contacts is >2.0
Ohms"
870 END IF
880 NEXT I
Summary of Contents for VXI Bus 75000 C Series
Page 1: ...Keysight E1463A 32 Channel 5 Amp Form C Switch Service Manual 75000 Series C...
Page 2: ......
Page 6: ...6...
Page 61: ......