![Keysight VXI Bus 75000 C Series Service Manual Download Page 24](http://html1.mh-extra.com/html/keysight/vxi-bus-75000-c-series/vxi-bus-75000-c-series_service-manual_1961145024.webp)
26
Keysight E1463A 32-Channel Form C Switch Service Guide
Verification Tests
Performance Verification
Channel 0 C to NC Measurements
1
Make Hardware Connections
–
Turn mainframe power OFF
–
Set DMM to measure 4-wire Ohms
–
Connect DMM leads as shown in Figure 2-3
–
Turn mainframe power ON
2
Measure Channel 0 C to NC Resistance
–
Send *RST to Module
–
Send CLOS (@nn00:nn31) to close all channels, where nn = card #
(typically 01)
–
Send OPEN (@nn00) to open channel 0, where nn = card # (typically 01)
–
Trigger the DMM with TRIG SGL and note reading
–
Enter the result in Table 2-1 for Channel 0 C to NC
Figure 1-5
Common to NO Test Connections
Summary of Contents for VXI Bus 75000 C Series
Page 1: ...Keysight E1463A 32 Channel 5 Amp Form C Switch Service Manual 75000 Series C...
Page 2: ......
Page 6: ...6...
Page 61: ......