Appendixes
160
Keysight U8903A User’s Guide
You may configure the arbitrary file using the format as shown in the example
below. The file is saved in the *.arb format.
#Vpeak: 2
#DC offset: 0
#Points:
0
–0.2
–0.4
.
.
.
The following table shows the allowable range of values for each arbitrary file
parameter.
The sampling rate for the arbitrary waveform is fixed at 312.5 kHz. Thus, the
interval between samples is 3.2
μ
s
(
1/312.5).
Parameter
Range
Vpeak
– 0 to 22.6 Vp (Balanced output connection)
– 0 to 11.3 Vp (Unbalanced or Common output connection)
DC offset
–11.3 V to 11.3 V
Points
32 to 32768 points
NOTE
When the DC offset and amplitude are added together, it must not exceed the
maximum voltage for the current output connection type:
– For the Balanced output connection, Vpeak + |DC offset| must be within 0 V
and 22.6 V.
– For the Unbalanced and Common mode output connections,
Vpeak + |DC offset| must be within 0 V and 11.3 V.