2
Digital Multimeter Operation
Making Measurements
38
U3606A User’s and Service Guide
Performing frequency, pulse width, and duty cycle
measurements
There are two measuring paths for the frequency, pulse width, and duty
cycle measurements — voltage or current. Therefore, before setting the
frequency measurement, it is advisable to configure the AC voltage (see
“Measuring AC voltage” on page 27
) or AC current (see
) measurements first.
Making the connections
Connect the test leads as shown below:
Figure 2-5
Terminal connections for frequency, pulse width, and duty cycle
measurements via the voltage path