Digital Multimeter Operation
2
Making Measurements
U3606A User’s and Service Guide
27
Measuring AC voltage
1
Press
again (until the AC annunciator is shown on the display) to
make AC voltage measurements.
2
Connect the red and black test leads to input terminals
V
(red) and
LO
(black) respectively as shown in
.
3
Probe the test points and read the display.
Table 2-2
AC voltage measurement summary
Item
Description
Available ranges
100.000 mV, 1.00000 V, 10.0000 V, 100.000 V, 750.00 V
Measurement method
AC coupled true rms
Input impedance
1 M
Ω
± 2% range (typical) in parallel with capacitance < 120 pF
Input protection
750 V
rms
on all ranges
Crest factor
For < 5:1 errors included. Limited by the peak input and 100 kHz
bandwidth. Maximum 3.0 at full scale.
Peak input
300% of range. Limited by maximum input.
V
AC annunciator