Digital Multimeter Operation
2
Making Measurements
U3606A User’s and Service Guide
31
Measuring AC (rms) current
1
Press
again (until the AC annunciator is shown on the display) to
make AC current measurements.
2
Connect the red and black test leads to input terminals
I
(red) and
LO
(black) respectively as shown in
.
3
Probe the test points and read the display.
Table 2-4
AC current measurement summary
Item
Description
Available ranges
10.0000 mA, 100.000 mA, 1.00000 A, 3.0000 A
Measurement method
AC coupled true rms
Burden voltage and shunt
resistance
•
< 0.2 V/10
Ω
for 10 mA range
•
< 0.2 V/1
Ω
for 100 mA range
•
< 0.3 V/0.1
Ω
for 1 A range
•
< 0.7 V/0.01
Ω
for 3 A range
Input protection
Protected with a 3.15 A/500 V, FF fuse
Crest factor
For < 5:1 errors included. Limited by the peak input and 100 kHz
bandwidth. Maximum 3.0 at full scale.
Peak input
300% of range. Limited by maximum input.
I
AC annunciator