Service and Maintenance
4
Keysight U1701B User’s and Service Guide
63
Specification Validation
You can perform self-validation of the capacitance meter's accuracy by using the
recommended equipment with the specified test ranges below.
Table 4-2
Recommended Equipment List
Table 4-3
Ranges of Functional Validation
Standard Source
Operating Range
Limit
Recommended Equipment
Capacitance
Calibrator
1 nF ~10 nF
100 nF ~10 mF
± 0.5%
± 0.25%
Fluke 5520A
or equivalent
Range
Test Value Used
1000.0 p
500 p
1000.0 n
500 n
1000.0 µ
500 µ
199.99 m
100 m