2
Features and Functions
30
Keysight U1701B User’s and Service Guide
Tolerance Mode
The tolerance mode has
1%
,
5%
,
10%
, and
20%
tolerance ranges. To enter the
tolerance mode, insert a standard value into the socket. Press the
TOL
key to set
the display value as the standard reference. Similarly, the DH value which appears
on the primary display can be used as a standard value to sort the components.
Press
TOL
key again to cycle through
1%
,
5%
,
10%
and
20%
tolerance and select
the desired tolerance. The meter range locks as tolerance mode.
This mode can not be set under following condition:
– After setting the recording mode.
– After setting HI/LO Audible Alert mode.
– The tested display is either
OL
or below 10 count.
This function is designed for sorting the values. The beeper will beep three times
when the reading is out of the selected tolerance. A single beep represents that
the reading is within the selected tolerance.
NOTE
To exit tolerance mode, press and hold
TOL
key for more than 1 s