Demonstrations

Demonstration 1: 

Getting started

The signal under test is simply a 

continuous-wave (CW) RF signal from 

the MXG. Recommended settings 

for signal generator are 1 GHz center 

frequency and –10 dBm RF output 

level.

Instructions  

Keystrokes

On the MXG signal generator

Preset the MXG

[Preset]

Set up the signal under test at 1 GHz 

center frequency and –10 dBm output 

level

[FREQ] [1] {GHz}; [AMPLD] [-] [10] {dBm}

Turn off modulation

[Mod On/Off] and ensure the LED under the key 

is off

Turn on RF output

[RF On/Off] and ensure the LED under the key  

is lit

On the X-Series signal analyzer

Switch to Phase Noise mode

[Mode] {Phase Noise}

Set the signal analyzer with a carrier 

frequency of 1 GHz and an amplitude 

reference level of 0 dBm

[FREQ Channel] {Carrie Freq} [1]{GHz}; [AMPTD 

Y Scale] {Ref Value} [0] {dBm}

View the signal at spectrum domain 

using “Monitor Spectrum”. It is 

always a good idea to use the moni-

tor spectrum measurement to make 

sure your X-Series signal analyzer set 

to the right frequency.

[Meas] {Monitor Spectrum}

Figure 2. Monitor spectrum to check the signal coming from the DUT

Summary of Contents for N9068A

Page 1: ...s signal analyzers with the N9068A W9068A phase noise measurement application Each demonstration is given a brief description of its function and the corresponding measurement steps on the signal generator and or signal analyzer Keysight Technologies N9068A W9068A Phase Noise X Series Measurement Application Demo Guide ...

Page 2: ...PXA N9020A MXA N9010A EXA or N9000A CXA Option 503 or any other frequency option Phase noise measurement application N9068A PXA MXA EXA W9068A CXA Option 2FP fixed perpetual license Option 2TP transportable perpetual license not available for CXA Figure 1 Demonstration setup Demonstration Setup Connect the X Series and MXG Perform the following steps to interconnect the X Series and MXG see Figure...

Page 3: ...f and ensure the LED under the key is off Turn on RF output RF On Off and ensure the LED under the key is lit On the X Series signal analyzer Switch to Phase Noise mode Mode Phase Noise Set the signal analyzer with a carrier frequency of 1 GHz and an amplitude reference level of 0 dBm FREQ Channel Carrie Freq 1 GHz AMPTD Y Scale Ref Value 0 dBm View the signal at spectrum domain using Monitor Spec...

Page 4: ...ength and view it superimposed with the raw data trace Figure 3 Log plot phase noise with a smoothed trace and decade table turned on Instructions Keystrokes On the X Series in Phase Noise mode View log plot Meas Log Plot Set the signal analyzer with a carrier fre quency of 1 GHz and an amplitude reference level of 0 dBm FREQ Channel Carrie Freq 1 GHz AMPTD Y Scale Ref Value 0 dBm Set the start of...

Page 5: ...as Log Plot Set the signal analyzer with a carrier frequency of 1 GHz and an amplitude reference level of 0 dBm FREQ Channel Carrie Freq 1 GHz AMPTD Y Scale Ref Value 0 dBm Set the start offset to 100 Hz and stop offset to 100 MHz and restart the log plot measurement SPAN X Scale Start Offset 100 Hz Stop Offset 100 MHz then Restart Activate markers for the rms jitter measurements Marker Function R...

Page 6: ...quency to 10 kHz Trace Detector Ref Trace On On the MXG Set the incremental frequency at 10 Hz per step FREQ Incr Set 10 kHz then or to manually adjust the carrier frequency drift by 10 kHz step On the X Series Phase Noise mode Adjust the signal analyzer s Y axis scale according to the 10 Hz step set on the MXG Manually adjust the RF carrier frequency from the output of the signal generator up or ...

Page 7: ... Keystrokes On the X Series Phase Noise mode Start DANL measurement and set the average number to 20 The signal analyzer averages 20 measurements wait for the measurement to finish the Avg Hold counter in the measurement status bar at the top of the display helps keep track of the status Mode preset Meas Log Plot Meas Setup Avg Number On 20 Enter Meas Type DANL Restart Save DANL as the reference t...

Page 8: ...lectronic Measurement Group DEKRA Certified ISO 9001 2008 Quality Management System Keysight Channel Partners www keysight com find channelpartners Get the best of both worlds Keysight s measurement expertise and product breadth combined with channel partner convenience wwww keysight com find n9068a This information is subject to change without notice Keysight Technologies 2011 2014 Published in U...

Reviews: