background image

7

Demonstration 5: 

Displayed average noise 
level (DANL) measurements 
and reference trace  
subtractions

The DANL floor of the signal analyzer 

may adversely affect the phase noise 

measurement results, particularly  

at the wider offset. The DANL mea-

surement provided by the N9068A/

W9068A measures and references the 

X-Series signal analyzer to the carrier 

amplitude.  

Furthermore, the measured DANL 

can be used as a reference trace 

when implementing reference trace 

subtraction to eliminate the analyzer’s 

broadband noise effect from the 

phase noise measurements.

Instructions

Keystrokes

On the X-Series Phase Noise mode:

Start DANL measurement and set the 

average number to 20. The signal analyzer 

averages 20 measurements – wait for 

the measurement to finish (the Avg|Hold 

counter in the measurement status bar at 

the top of the display helps keep track of 

the status).

[Mode preset] [Meas] {Log Plot} [Meas 

Setup] {Avg Number On} [20] {Enter}, {Meas 

Type} {DANL} [Restart]

Save DANL as the reference trace for the 

following DANL subtraction

[Trace/Detector] {Select Trace} {Trace 2}  

{More 1 of 2} {Copy/Exchange} {From Trace}  

{Trace 2} {To Trace} {Trace 3} {Copy Now}

Implement DANL subtraction

[Meas Setup] {More 1 of 2} {Cancellation} 

{Cancellation On} {Ref Trace} {Trace 3}; 

[Meas Setup] {Meas Type} {Phase Noise} 

[Single]

Figure 6. DANL measurement result

Figure 7. Phase noise plot with DANL subtraction

Summary of Contents for N9068A

Page 1: ...s signal analyzers with the N9068A W9068A phase noise measurement application Each demonstration is given a brief description of its function and the corresponding measurement steps on the signal generator and or signal analyzer Keysight Technologies N9068A W9068A Phase Noise X Series Measurement Application Demo Guide ...

Page 2: ...PXA N9020A MXA N9010A EXA or N9000A CXA Option 503 or any other frequency option Phase noise measurement application N9068A PXA MXA EXA W9068A CXA Option 2FP fixed perpetual license Option 2TP transportable perpetual license not available for CXA Figure 1 Demonstration setup Demonstration Setup Connect the X Series and MXG Perform the following steps to interconnect the X Series and MXG see Figure...

Page 3: ...f and ensure the LED under the key is off Turn on RF output RF On Off and ensure the LED under the key is lit On the X Series signal analyzer Switch to Phase Noise mode Mode Phase Noise Set the signal analyzer with a carrier frequency of 1 GHz and an amplitude reference level of 0 dBm FREQ Channel Carrie Freq 1 GHz AMPTD Y Scale Ref Value 0 dBm View the signal at spectrum domain using Monitor Spec...

Page 4: ...ength and view it superimposed with the raw data trace Figure 3 Log plot phase noise with a smoothed trace and decade table turned on Instructions Keystrokes On the X Series in Phase Noise mode View log plot Meas Log Plot Set the signal analyzer with a carrier fre quency of 1 GHz and an amplitude reference level of 0 dBm FREQ Channel Carrie Freq 1 GHz AMPTD Y Scale Ref Value 0 dBm Set the start of...

Page 5: ...as Log Plot Set the signal analyzer with a carrier frequency of 1 GHz and an amplitude reference level of 0 dBm FREQ Channel Carrie Freq 1 GHz AMPTD Y Scale Ref Value 0 dBm Set the start offset to 100 Hz and stop offset to 100 MHz and restart the log plot measurement SPAN X Scale Start Offset 100 Hz Stop Offset 100 MHz then Restart Activate markers for the rms jitter measurements Marker Function R...

Page 6: ...quency to 10 kHz Trace Detector Ref Trace On On the MXG Set the incremental frequency at 10 Hz per step FREQ Incr Set 10 kHz then or to manually adjust the carrier frequency drift by 10 kHz step On the X Series Phase Noise mode Adjust the signal analyzer s Y axis scale according to the 10 Hz step set on the MXG Manually adjust the RF carrier frequency from the output of the signal generator up or ...

Page 7: ... Keystrokes On the X Series Phase Noise mode Start DANL measurement and set the average number to 20 The signal analyzer averages 20 measurements wait for the measurement to finish the Avg Hold counter in the measurement status bar at the top of the display helps keep track of the status Mode preset Meas Log Plot Meas Setup Avg Number On 20 Enter Meas Type DANL Restart Save DANL as the reference t...

Page 8: ...lectronic Measurement Group DEKRA Certified ISO 9001 2008 Quality Management System Keysight Channel Partners www keysight com find channelpartners Get the best of both worlds Keysight s measurement expertise and product breadth combined with channel partner convenience wwww keysight com find n9068a This information is subject to change without notice Keysight Technologies 2011 2014 Published in U...

Reviews: