Installation Note N9030-90111

15

Option 107, Audio Input and Digitizer Upgrade Kit

Verify the Installation

1.

Turn on the analyzer and wait for the instrument to fully boot up. 

2.

Verify the spectrum analyzer application loads and sweeps as expected. Press 

System

Show 

Hardware

 on the analyzer and verify that the Audio Digitizer identifies itself as:

Audio Input 100 K Ohms

3.

Press 

System

Show System

 on the analyzer to display a list of all installed options. You 

should see N9030B-107.

Summary of Contents for N9030B

Page 1: ...Keysight Technologies N9030B Signal Analyzer Option 107 Audio Input and Digitizer Upgrade Kit Installation Note ...

Page 2: ...mplied warranties of merchantability and fitness for a particular purpose Keysight Technologies shall not be liable for errors contained herein or for incidental or consequential damages in connection with the furnishing performance or use of this material Manual Part Number N9030 90111 Edition Edition 2 October 2018 Supersedes none Printed in USA Malaysia Published by Keysight Technologies Inc 14...

Page 3: ...adjustments and performance verification tests However the completion of these tests does not guarantee that the instrument meets all advertised specifications Software and test equipment is required for making adjustments and for performance verification testing Information on how to obtain this software can be found at www keysight com find calibrationsoftware While Keysight does recommend that ...

Page 4: ...ysis Bandwidth N9030B BBA I Q Baseband Inputs 2 This upgrade kit installation requires that you re install instrument software The latest revision of the PXA Signal Analyzer software may be downloaded from http www keysight com find xseries_software The instrument software must be installed AFTER the hardware in the kit is installed Follow the procedure in this document for redeeming and installin...

Page 5: ...crew FLT HD Torx T10 M3 x 0 5 6 mm long 0515 1227 1 Lock Washer Internal tooth 2190 0068 1 Hex Nut 15 32 32 Interface Board 2950 0035 1 Coaxial Cable Assembly w Toroids N9020 60344 1 Cover Plate Audio N9030 00004 1 PCA Audio Digitizer N9030 60047 1 Overlay Audio Input N9030 80062 1 Installation Note this note 1 Option Upgrade Entitlement Certificate n a ...

Page 6: ...you disassemble the instrument turn the power switch to Standby wait until the instrument has completely shut down and unplug the instrument Failure to unplug the instrument can result in personal injury Electrostatic discharge ESD can damage or destroy electronic components All work on electronic assemblies should be performed at a static safe workstation Refer to the documentation that pertains ...

Page 7: ...es 2 on each side of the instrument 3 Remove the four key locks from the instrument four bottom feet 4 Remove the four instrument bottom feet 5 Using the T 20 driver remove the four screws including washers 3 that hold the rear feet 4 in place 6 Pull the instrument cover 5 off towards the rear of the instrument If the instrument is placed on its face during any of the following procedures be sure ...

Page 8: ...8 Installation Note N9030 90111 Option 107 Audio Input and Digitizer Upgrade Kit Figure 1 Instrument Outer Case Removal ...

Page 9: ...emoval 1 Refer to Figure 2 To remove the top brace 1 use the T 10 driver to remove the twelve panhead screws 3 0515 0372 attaching the top brace to the chassis Remove the twelve screws fourteen screws with Option B1X 4 0515 1946 attaching the top brace to the boards Figure 2 Top Brace Removal ...

Page 10: ...te N9030 90111 Option 107 Audio Input and Digitizer Upgrade Kit Option 107 Hardware Installation 1 Refer to Figure 3 Locate and install the N9030 60047 Audio Digitizer PCA into slot 7 Figure 3 Option 107 Audio Digitizer ...

Page 11: ...anel overlay Figure 4 Cover Plate 3 Replace the cover plate with the one provided with this kit N9030 00004 4 Secure the new cover plate with the two screws and torque to 9 in lbs Figure 4 5 Locate the new audio input coaxial cable provided with this kit N9020 60344 6 Insert the BNC end of the cable into the back side of the front frame assembly so that it comes out the top hole of the newly insta...

Page 12: ...off layer from the back of the overlay Carefully locate and center the new Audio Input Overlay over the BNC connector and within the front frame casting relief 10 Route the cable between the inner and outer chassis brackets above the fans and then connect to P6 on the Audio Digitizer assembly Top Brace Replacement 1 Refer to Figure 2 Replace the top brace 1 using the 0515 1946 screws There is an a...

Page 13: ... http www keysight com find pxa_software This will guarantee The analyzer has the latest instrument software version to support Option 107 FPGA updates are done to the Audio Digitizer hardware The drivers are installed for the new hardware Licensing the Option Installation Procedure over USB 1 Locate the Option Upgrade Entitlement Certificate from the kit 2 Redeem the Option Upgrade Entitlement Ce...

Page 14: ...a few minutes When the License File is consumed the Keysight License Manager will display a Successful License Installation message similar to the one shown in Figure 7 Figure 7 Successful License Installation Alternate Installation Procedure The License File can be manually installed over USB or LAN by placing the license file in the following folder on the PXA C Program Files Agilent licensing ...

Page 15: ...or the instrument to fully boot up 2 Verify the spectrum analyzer application loads and sweeps as expected Press System Show Hardware on the analyzer and verify that the Audio Digitizer identifies itself as Audio Input 100 K Ohms 3 Press System Show System on the analyzer to display a list of all installed options You should see N9030B 107 ...

Page 16: ...udio Amplitude Accuracy Option 107 Audio Distortion Option 107 IMPORTANT A full calibration is required to assure the instrument meets all specifications The end user must ultimately determine whether they want a full calibration to be performed after the installation of this upgrade or not If a full calibration is required arrangements regarding the level of calibration must be made between the e...

Page 17: ...This information is subject to change without notice Keysight Technologies 2018 Edition 2 October 2018 N9030 90111 www keysight com ...

Reviews: