Installation Note N5247-90116 41
Description of the Upgrade
Installation Procedure for the Upgrade
Step 23. Verify the PNA Analyzer Program is Running with the Correct
Options
1.
Start the Network Analyzer program.
2.
Once the Network Analyzer program is running:
—
Press
Help
>
About NA
and verify that Option 205 is listed in the
PNA application.
Step 24. Perform Post-Upgrade Adjustments and Calibration
Adjustments
The following adjustments must be made due to the hardware changes of the
analyzer.
— Default EE adjustment - select the LO Drive-NF adjustment and either
adjust or initialize the values.
— Source Adjustment
— IF Gain Adjustment
— Receiver Characterization
— Receiver Adjustment
— IF Response Adjustment (N5247A with Option 090, 093, 094, or N5247B
with Option S93090xA, S93093A, or S93094A Only)
— Noise Figure Adjustment (Option N5247A 029 or N5247B 029 with
S93029A Only)
These adjustments are described in the PNA Service Guide and in the PNA
on-line HELP. A list of equipment required to perform these adjustments is also
found in the service guide.
To view this service guide information, click the Chapter 3 bookmark “Tests
and Adjustments” in the PDF Service Guide
1
.
After the specified adjustments have been performed, the analyzer should
operate and phase lock over its entire frequency range.
If if the option(s) have not been enabled or if your older options have not
been removed, contact Keysight Technologies. Refer to
Assistance from Keysight” on page 6
.