Installation Note N5245-90125                                                                                                                                  7

Installation Procedure for the Upgrade

Step 1. Remove the Top Shield, Gasket, and Old Polyiron

1.

Place the noise figure board assembly so that the top shield is facing up 
(item 

). Refer to 

Figure 1 on page 8

2.

Remove the twenty screws 0515-0665 (item 

) from the top shield. 

Inspect the screws and discard any that are damaged or unusable and 

replace with the spare 0515-0665 screws.

3.

Remove the old top shield (item 

) from the noise board (item 

) and 

bottom shield (item 

) assembly. Refer to 

Figure 1 on page 8

.

4.

Remove the gasket from the old top shield (seven places) (items [G1] to 

[G7]) using a mini/precision flathead screw driver. Save gasket for reuse. 
Refer to 

Figure 2 on page 9

 and 

Figure 3 on page 9

.

When removing the gasket, be careful not to damage the gasket material.

When removing the gasket, be careful not to damage/stretch the gasket 

material.

Summary of Contents for N5245-60125

Page 1: ...n Note Keysight Noise Figure Board No Tabs Revision Kit For PNA models with serial number prefixes 5200 To convert a N5245 60124 with tabs board to a N5245 60124 no tabs board Keysight Kit Number N524...

Page 2: ...U S government customers under its standard commercial license which is embodied in its End User License Agreement EULA a copy of which can be found at http www keysight com find sweula The license s...

Page 3: ...he work or need advice contact Keysight Contacting Keysight Assistance with test and measurements needs and information on finding a local Keysight office are available on the Web at http www keysight...

Page 4: ...odel number use the following steps 1 Go to www keysight com 2 In the Search box enter the model number of the analyzer Ex N5245B and click Search 3 Click Technical Support Manuals 4 Click Service Man...

Page 5: ...ESD grounding wrist strap 9300 1367 5 ft grounding cord for wrist strap 9300 0980 2 x 4 ft conductive table mat and 15 ft grounding wire 9300 0797 ESD heel strap for use with conductive floors 9300 13...

Page 6: ...2 Reinstall the Gasket x7 and Add New Polyiron to the New Top Shield Step 3 Reinstall the Top Shield Table 1 Contents of Upgrade Kit N5245 60125 Ref Desig Description Qty Part Number Installation not...

Page 7: ...that are damaged or unusable and replace with the spare 0515 0665 screws 3 Remove the old top shield item from the noise board item and bottom shield item assembly Refer to Figure 1 on page 8 4 Remove...

Page 8: ...8 Installation Note N5245 90125 Installation Procedure for the Upgrade Figure 1 Noise Figure Board Assembly...

Page 9: ...5 90125 9 Installation Procedure for the Upgrade Figure 2 Remove Gasket from Top Shield Use Caution During Removal Figure 3 Remove Gasket from Top Shield Using Mini Precision Flathead Screwdriver Use...

Page 10: ...4 3 Take the new magnetic absorber plate polyiron similar to item in Figure 2 on page 9 and remove the adhesive paper Take the new shield N5245 20190 item and install the magnetic absorber plate polyi...

Page 11: ...Installation Note N5245 90125 11 Installation Procedure for the Upgrade Figure 5 Install Reuse Gasket and Using Pinch Pad Points Use Caution During Removal Examples of Pinch Pad Points...

Page 12: ...he Upgrade Step 3 Reinstall the Top Shield 1 Install the new top shield onto the noise figure board item and bottom shield item 2 Reinstall the twenty 0515 0665 screws using the T 10 and torque to 9 i...

Page 13: ...Installation Note Xxxxx xxxxx 3...

Page 14: ...This information is subject to change without notice Keysight Technologies 2012 2017 Edition 1 January 2017 www keysight com...

Reviews: