N4965A Getting Started Guide
12
Multi-Channel BERT 12.5 Gb/s User Guide
1.6
Unpacking
Carefully remove the instrument from the case in an ESD-safe environment.
1.7
Important Notes
Use ESD protection at all times when using the instrument.
Review min/max specifications before applying input signals.
Use high quality SMA-connectors on the SMA ports.
Excessive mating of low quality SMA components to 2.92 mm female
receptacles may degrade the 2.92 mm female receptacle.
Leave dust covers on unused connectors.
Situate the instrument away from heat sources, do not block the fan, and do not
block the exhaust vent (minimum of 8 cm/3 inches clearance).
When using multiple remote heads, allow adequate space around heads for
cooling. If heads are stacked then additional cooling, e.g. using a fan, may be
required.
Use 50 Ω terminations on all unused differential ports.
Use a 8 lbf-in (90 N-cm) torque wrench when attaching connectors.
1.8
Measurement Best Practices
When using differential-mode connections, ensure the cables are phase
balanced for best performance.
Differential connectors may be used single-ended if unused outputs are
terminated in 50 Ω.
Use high quality cables and connector savers (or adaptors).
Keep cable lengths short and minimize the number of cable bends.
Use a 8 lb-in (90 N-cm) torque wrench when attaching connectors.