74
Keysight N4372E Lightwave Component Analyzer User’s Guide
4
Making Measurements
a
If you want to use a different electrical calibration for the user
calibration measurement and for the DUT measurement, "Electrical
Calibration" can be used as the user calibration measurement in the
"CalSet for User Calibration" combo box.
This electrical calibration will only be used for the “user calibration”
measurements. Afterwards the LCA returns to the currently
activated electrical calibration for the DUT measurements.
The currently activated electrical calibration is preselected in the
"CalSet for User Calibration" combo box. This preselection is fine
and you have nothing to do if the same electrical calibration is used
for the user calibration measurement and for the DUT measurement.
b
Select “Reduce IF-BW for user calibration measurement” to reduce
noise on the user calibration data. The default bandwidth reduction
factor is defined in the LCAConfig.xml file.
c
Click
to proceed.
d
Choose whether you want to save the calibration data for later use.