Keysight N4372E Lightwave Component Analyzer User’s Guide
53
Making Measurements
4
Electrical calibration for differential EO measurements
Calibrate before you start using the Lightwave Component Analyzer
application. Alternatively, return to the EE measurement mode, as
described in
Returning to Electrical Measurements
calibrate. For calibrating the Lightwave Component Analyzer using an
automated electrical calibration module, follow these steps:
1 Set your measurement parameters.
2 From the “Calibration” menu, select “Smart Cal”.
3 Select “3 Port Cal”.
4 Select 3 Port Cal Configuration according to the planned port
configuration for EO measurement and click the “Next” button.
Figure 6
Port-configuration for the 3 port calibration
5 Select DUT connectors and the Cal Kits (e.g. “1.0mm female” and
“85059B”).
6 Follow the instructions for further steps shown on the Network
Analyzer.
7 After the calibration has been completed, save the calibration file for
later reuse.