Performance Verification
7
Average Gain of Zoomed-In Channel
N2820/1A User’s Guide
97
Average Gain of Zoomed-In Channel
This procedure applies to both N2820A and N2821A probes.
1
Snap an N2825A head onto the N2820/1A probe and connect the
test equipment as shown in
. Be sure to connect the
positive side of the DC supply to the red lead on the N2825A head.
Figure 49
Test Setup for Measuring V
in
2
Set the DC power supply to +20 mV.
NOTE
Allow the probe to warm up for at least 20 minutes.
NOTE
Although this is the recommended setting, you can use any voltages up
to +35 mV.