Making Measurements
2
Burden Voltage of N2822/4A Probe Heads
N2820/1A User’s Guide
47
Burden Voltage of N2822/4A Probe Heads
When an N2822/4A head is used, the resistance of the head’s
connectors, sensor leads, and internal
R
SENSE
resistor is connected
in series with the DUT’s circuit and current is flowing through the
probe head, as shown in
Therefore, a burden voltage (IR) is introduced into your circuit.
Burden voltages may affect the circuit and will tend to increase the
noise. Burden voltages do
not
affect the measurement.
page 11 lists the resistances added by various probing
components.
Figure 20
V
burden
Developed Across Probing Components
V
burden
I
measured
R
shunt
×
=
NOTE
N2825A heads do not contribute a burden voltage as the current is
not
flowing through the head.