27
Keysight N2795/6A Single-Ended Probes
User’s Guide
2 Characteristics and
Specifications
This chapter provides the characteristics and specifications for the
N2795/6A probes. The probe and oscilloscope should be warmed
up for at least 20 minutes before any testing and the environmental
conditions should not exceed the probe’s specified limits.
Table 5
Environmental Specifications
Parameter
Characteristic
Operating Temperature
0°C to +50°C
Non-Operating Temperature
–40°C to +70°C (–40°F to +158°F)
Altitude
Operating; 4,000m (13,000 ft)
Operating Humidity
95% room humidity at +40°C
Non-Operating Humidity
90% room humidity at +65°C
ESD
8 kV HBM
Use
indoor
Table 6
Electrical Characteristics and Specifications (Sheet 1 of 2)
Item
N2795A
N2796A
Attenuation Ratio (at DC)
10:1 ± 0.5%
10:1 ± 0.5%
Bandwidth (–3 dB)
a
1 GHz
2 GHz
Probe Risetime (calculated, 10% – 90%)
350 ps
175 ps
Input Dynamic Range
–8V to +8V (DC or Peak AC)
–8V to +8V (DC or Peak AC)