Performance Verification
4
N2795/6A User’s Guide
41
12
Perform a Calibration for the
PORT 1
side of the Reference plane.
a
Press the
reflection
screen key.
b
Connect the open end of 85052D Calibration Standard
to the
non-pincher side of the PV/DS test board.
c
Select the
open
screen key under the
Forward
group.
d
Wait until the VNA beeps indicating that it has completed the
task.
e
Connect short end of Calibration Standard to the
non-pincher side of the PV/DS test board.
f
Select
short
screen key under the
Forward
group.
g
Wait until the VNA beeps indicating that it has completed the
task.
h
Connect load end of Calibration Standard to the non-pincher
side of the PV/DS test board.
i
Select the
loads
screen key under the
Forward
group.
j
Press
broadband
screen key selection.
k
Wait until the VNA beeps indicating that it has completed the
task.
l
Press the
done loads
screen key.
m
You have just calibrated one side of the reference plane.
13
Connect the other high quality SMA cable to the VNA’s
PORT 2
.