129
OIF CEI-56G LR PAM4 Calibrations
8
SJ Calibration
Overview
The SJ Calibration is performed to calibrate the Sinusoidal Jitter.
Connection Diagram
Connect the instruments as shown in
Figure 97
SJ Calibration connections for LR
ID
Parameters
Refer to “
Calibration Parameters in Debug Mode
Procedure
This calibration uses the Clock/8 pattern.
The Victim Generator PJ1 is set until the desired SJ value is measured.
Results
This calibration returns the results for the set PJ1 versus measured SJ, in
graphical and tabular formats.
Standard Name
Test ID
Long Reach
671101