Quick Reference
2
Keysight InfiniiVision 1000 X-Series Oscilloscopes User's Guide
61
Measurements
The
[Meas]
key lets you make automatic measurements on waveforms. Some
measurements can only be made on analog input channels.
The results of the most recent measurements are displayed in the Measurements
information area on the right-hand side of the screen.
Cursors are turned on to show the portion of the waveform being measured for the
most recently selected measurement (bottom-most on the right-side
measurement are).
The units of math waveforms are described in
All measurements are available for analog channel waveforms. All measurements
except Counter are available for math waveforms other than FFT. A limited set of
measurements is available for math FFT waveforms. Use the cursors to make other
measurements on FFT.
NOTE
If a portion of the waveform required for a measurement is not displayed or does not display
enough resolution to make the measurement (approximately 4% of full scale), the result will
be displayed as greater than a value, less than a value, not enough edges, not enough
amplitude (low signal), or waveform is clipped.
NOTE
Post Acquisition Processing
In addition to changing display parameters after the acquisition, you can perform all of the
measurements and math functions after the acquisition. Measurements and math functions
will be recalculated as you pan and zoom and turn channels on and off. As you zoom in and out
on a signal using the horizontal scale knob and vertical volts/division knob, you affect the
resolution of the display. Because measurements and math functions are performed on
displayed data, you affect the resolution of functions and measurements.