background image

2

  

                                                                                                                            

  

Installation Note E8362-90004

Notice

THE MATERIAL CONTAINED IN THIS DOCUMENT IS PROVIDED “AS IS,” AND IS SUBJECT 
TO BEING CHANGED, WITHOUT NOTICE, IN FUTURE EDITIONS. FURTHER, TO THE 
MAXIMUM EXTENT PERMITTED BY APPLICABLE LAW, AGILENT DISCLAIMS ALL 
WARRANTIES, EITHER EXPRESS OR IMPLIED WITH REGARD TO THIS MANUAL AND ANY 
INFORMATION CONTAINED HEREIN, INCLUDING BUT NOT LIMITED TO THE IMPLIED 
WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE. 
AGILENT SHALL NOT BE LIABLE FOR ERRORS OR FOR INCIDENTAL 
OR CONSEQUENTIAL DAMAGES IN CONNECTION WITH THE FURNISHING, USE, OR 
PERFORMANCE OF THIS DOCUMENT OR ANY INFORMATION CONTAINED HEREIN. 
SHOULD AGILENT AND THE USER HAVE A SEPARATE WRITTEN AGREEMENT WITH 
WARRANTY TERMS COVERING THE MATERIAL IN THIS DOCUMENT THAT CONFLICT 
WITH THESE TERMS, THE WARRANTY TERMS IN THE SEPARATE AGREEMENT WILL 
CONTROL.

Getting Assistance from Agilent

By internet, phone, or fax, get assistance with all your test and measurement needs.

Online assistance:

www.agilent.com/find/assist

United States

(tel)

 1 800 452 4844

Latin America

(tel)

 (305) 269 7500

(fax)

 (305) 269 7599

Canada

(tel)

 1 877 894 4414

(fax)

 (905) 282-6495

Europe

(tel)

 (+31) 20 547 2323

(fax)

 (+31) 20 547 2390

Australia

(tel)

 1 800 629 485

(fax) (+61) 3 9210 5947

New

 

Zealand

(tel)

 0 800 738 378

(fax)

 (+64) 4 495 8950

Japan

(tel) (+81) 426 56 7832
(fax) (+81) 426 56 7840

Singapore

(tel)

 1 800 375 8100

(fax)

 (65) 836 0252

Malaysia

(tel) 

1 800 828 848

(fax)

 1 800 801 664

India

(tel)

 1 600 11 2929

(fax)

 000 800 650 1101

Hong Kong

(tel) 

800 930 871

(fax)

 (852) 2506 9233

Taiwan

(tel

) 0800 047 866

(fax)

 (886) 2 25456723

Philippines

(tel

) (632) 8426802

(tel) (PLDT subscriber only)

1 800 16510170

(fax)

 (632) 8426809

(fax) (PLDT subscriber only)

1 800 16510288 

Thailand

(tel) (outside Bangkok)

(088) 226 008

(tel) (within Bangkok)

(662) 661 3999

(fax)

 (66) 1 661 3714

People’s Republic of China

(tel) (preferred)

800 810 0189

(tel) (alternate

)

10800 650 0021

(fax)

 10800 650 0121

Summary of Contents for E8362-60109

Page 1: ...ogies Inc All rights reserved Installation Note Extended Frequency Range Upgrade Kit For E8362B PNA Series Microwave Network Analyzers WITHOUT the Configurable Test Set Option Network Analyzer Model N...

Page 2: ...GREEMENT WILL CONTROL Getting Assistance from Agilent By internet phone or fax get assistance with all your test and measurement needs Online assistance www agilent com find assist United States tel 1...

Page 3: ...ed to extend the frequency range of the E8362B PNA microwave network analyzer from 20 GHz to 40 GHz Products affected E8362B without Option 014 Configurable Test Set Installation to be performed by Ag...

Page 4: ...8 A26 Test port 2 coupler A27 Channel A first converter mixer 2 5087 7122 A30 Channel B first converter mixer N A Machine screw M3 0 x 18 for attaching A23 and A24 detectors 4 0515 0666 N A Machine sc...

Page 5: ...ption UNL ONLY W7 A23 channel R1 detector to A28 channel R1 mixer 1 E8364 20174 W8 A24 channel R2 detector to A29 channel R2 mixer 1 E8364 20175 Parts for analyzers WITH Options UNL and 016 ONLY W7 A2...

Page 6: ...ies within the network analyzer Static electricity can build up on your body and can easily damage sensitive internal circuit elements when discharged Static discharges too small to be felt can cause...

Page 7: ...ep 13 Replace the Front Panel Nameplate Step 14 Reinstall the Front Panel Assembly Step 15 Reinstall the Inner and Outer Covers Step 16 Enter the New Model Number Step 17 Verify that the New Model Num...

Page 8: ...r on the certificate matches the option described by this installation note If the model serial or option number do not match the expected results you will not be able to install the option If this is...

Page 9: ...amage the front panel assemblies 1 Disconnect the power cord if it has not already been disconnected 2 Using a T 20 TORX driver remove the strap handles item by loosening the screws item on both ends...

Page 10: ...ation Note E8362 90004 Step 3 Remove the Inner Cover 1 Place the analyzer top side up on a flat surface 2 Using a T 10 TORX driver remove the eleven screws item 3 Lift off the cover Figure 3 Inner Cov...

Page 11: ...om the sides of the frame CAUTION Before removing the front panel from the analyzer lift and support the front of the analyzer chassis 2 Slide the front panel over the test port connectors 3 Disconnec...

Page 12: ...5 Disconnect the A12 Source 20 to A22 Switch Splitter Cable at the A12 Source 20 1 Using a 5 16 inch wrench disconnect the A12 source 20 to A22 switch splitter cable W1 at the A12 source 20 as shown F...

Page 13: ...screws item that secure the receiver deck 3 Pull the latch pin item towards the opposite side of the analyzer to release the receiver deck 4 Lift the receiver deck to partially raise it then release t...

Page 14: ...move two RF cables item from the analyzer and retain for reinstallation later 7 Remove four screws item from the mixer bracket Retain the screws for reinstallation later 8 Remove the mixer bracket wit...

Page 15: ...Installation Note E8362 90004 15 Figure 7 Mixers Bracket Removal Figure 8 Mixers Replacement...

Page 16: ...RF cable item from each test port coupler 3 Using a T 10 TORX driver remove eight screws items through from the test set deck front panel to release it 4 Remove the test set deck front panel from the...

Page 17: ...e 9 2 Install a flange nut item removed from the old couplers onto each new coupler Using a 1 inch torque wrench tighten each flange nut to 72 in lbs 3 Reconnect the two cables item to the test port c...

Page 18: ...J22 P1 BIAS T and J25 P2 BIAS T and release them from the cable clamp item 3 Disconnect the ribbon cables at the step attenuators item 4 Remove three screws item from each bracket Retain the screws f...

Page 19: ...as Tees Replacement Reinstall the Detector Bias Tee Attenuator Brackets Refer to Figure 11 1 Place the brackets with the attenuators and bias tees attached into the analyzer as shown in Figure 11 2 Re...

Page 20: ...ect the wrapped wire control cables from the A23 and A24 detectors A23 detector to A16 test set motherboard J5 R1 DET A24 detector to A16 test set motherboard J6 R2 DET Install the A21 SOMA 50 Refer t...

Page 21: ...er to A23 channel R1 detector W4 A22 switch splitter to A24 channel R2 detector W37 A23 channel R1 detector to A16 test set motherboard J204 R1 W38 A24 channel R2 detector to A16 test set motherboard...

Page 22: ...lease the receiver deck 2 Lift the receiver deck to partially lower it then release the latch pin item Lower the receiver deck to its fully lowered position and ensure that the latch pin latches in th...

Page 23: ...A12 Source 20 to A21 SOMA 50 Cable at the A12 Source 20 1 Place the analyzer top side up on a flat surface 2 Using a 5 16 inch wrench connect the A12 source 20 to A21 SOMA 50 cable W1 at the A12 sour...

Page 24: ...ace can be cleaned with isopropol alcohol Install the New Nameplate IMPORTANT If the Option 050 upgrade kit will be installed to extend the frequency to 50 GHz do not install the new nameplate provide...

Page 25: ...the A3 front panel interface board 2 Slide the front panel over the test port connectors being careful to align the power switch and floppy disk drive to their corresponding front panel cutouts Ensure...

Page 26: ...l the Inner Cover 1 Place the inner cover on the analyzer as shown There are two alignment pins on the front frame that align with holes in the front of the cover to ensure proper alignment 2 Using a...

Page 27: ...its front panel without the handles This will damage the front panel assemblies 1 Slide the cover onto the frame 2 Using a T 20 TORX driver install the four rear panel feet item by installing the cent...

Page 28: ...ct click NO and repeat the repair 8 When the installation is complete click Exit Front Panel Keys Procedure 1 In the COMMAND block press Menu Dialog 2 In the NAVIGATION block press the right tab and a...

Page 29: ...OK 3 On the analyzer front panel press Preset Verify that the full frequency range of the analyzer is correct The analyzer should preset to a frequency range of 10 MHz to 40 GHz Front Panel Keys Proc...

Page 30: ...r should now operate and phase lock over its entire frequency range If you experience difficulty with the basic functioning of the analyzer contact Agilent Refer to Getting Assistance from Agilent on...

Reviews: