![Keysight E8362-60109 Installation Note Download Page 29](http://html1.mh-extra.com/html/keysight/e8362-60109/e8362-60109_installation-note_1961285029.webp)
Installation Note E8362-90004
29
Step 17. Verify that the New Model Number Is Shown Correctly
Procedure Requirements
• The analyzer must be powered up and operating to perform this procedure.
• The Network Analyzer program must be running.
• A mouse is recommended for this procedure but is not required.
Mouse Procedure
1. On the analyzer’s
Help
menu, click
About Network Analyzer
.
2. Verify that the new model number, E8363B, is shown correctly in the display. Click
OK
.
3. On the analyzer front panel, press
Preset
. Verify that the full frequency range of the
analyzer is correct. The analyzer should preset to a frequency range of 10 MHz to 40 GHz.
Front Panel Keys Procedure
1. In the
COMMAND
block, press
Menu/Dialog
.
2. In the
NAVIGATION
block, press the right tab and arrow keys to move over to the
Help
menu, and down to the
About Network Analyzer
selection. Press
Click
.
3. Verify that the new model number, E8363B, is shown correctly, then press
Click
.
4. On the analyzer front panel, press
Preset
. Verify that the full frequency range of the
analyzer is correct. The analyzer should preset to a frequency range of 10 MHz to 40 GHz.