170
Keysight InfiniiVision 3000T X-Series Oscilloscopes User's Guide
11
Triggers
Changes to the trigger setup are applied immediately. If the oscilloscope is
stopped when you change a trigger setup, the oscilloscope uses the new
specification when you press
[Run/Stop]
or
[Single]
. If the oscilloscope is running
when you change a trigger setup, it uses the new trigger definition when it starts
the next acquisition.
You can use the
[Force Trigger]
key to acquire and display data when triggers are
not occurring.
You can use the
[Mode/Coupling]
key to set options that affect all trigger types (see
, “Trigger Mode/Coupling,” starting on page 207).
You can save trigger setups along with the oscilloscope setup (see
“Save/Email/Recall (Setups, Screens, Data),” starting on page 321).
Triggers - General
Information
A triggered waveform is one in which the oscilloscope begins tracing (displaying)
the waveform, from the left side of the display to the right, each time a particular
trigger condition is met. This provides stable display of periodic signals such as
sine waves and square waves, as well as nonperiodic signals such as serial data
streams.
The figure below shows the conceptual representation of acquisition memory. You
can think of the trigger event as dividing acquisition memory into a pre-trigger
and post-trigger buffer. The position of the trigger event in acquisition memory is
defined by the time reference point and the delay (horizontal position) setting (see
“To adjust the horizontal delay (position)"
Adjusting the Trigger Level
You can adjust the trigger level for a selected analog channel by turning the
Trigger Level knob.
Pre-Trigger Buffer
Post-Trigger Buffer
Acquisition Memory
Trigger Event
Summary of Contents for DSOX3012T
Page 78: ...78 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 2 Horizontal Controls...
Page 88: ...88 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 3 Vertical Controls...
Page 98: ...98 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 4 FFT Spectral Analysis...
Page 148: ...148 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 7 Digital Channels...
Page 154: ...154 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 8 Serial Decode...
Page 160: ...160 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 9 Display Settings...
Page 232: ...232 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 13 Acquisition Control...
Page 242: ...242 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 14 Cursors...
Page 288: ...288 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 16 Mask Testing...
Page 320: ...320 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 19 Waveform Generator...
Page 340: ...340 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 21 Print Screens...
Page 374: ...374 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 23 Web Interface...
Page 394: ...394 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 24 Reference...