1 Introduction
CAUTION, VENTILATION REQUIREMENTS: When installing the instrument(s)
into a cabinet consideration shall be given to the convection flow into and out
of the cabinet. Consideration shall also be given to the individual instruments to
avoid having the heated discharge of one instrument, now becoming the
cooling intake air for another instrument.
Another area of concern is verification that the maximum ambient operating
temperature of the instrument(s) is not exceeded by cabinet installation.
Keysight recommends forced air convection whenever an instrument(s) are
installed in a cabinet and further recommends that the maximum operating
temperature of the cabinet be reduced 10°C from the lowest, of the maximum
operating temperature of a single instrument.
If there are any concerns or special requirements an Keysight Field Engineer
should be consulted to assure instrument(s) temperature compliance and
performance.
Install the instrument according to the enclosure protection provided. This
instrument does not protect against the ingress of water. This instrument
protects against finger access to hazardous parts within the enclosure.
Is your product software up-to-date? Keysight periodically releases software
updates to fix known defects and include product enhancements. To search for
software updates for your product, go to www.keysight.com/find/TechSupport.
16
N1000A Quick Start Guide
Summary of Contents for DCA-X Series
Page 1: ...Keysight N1000A DCA X Series Wide Bandwidth Oscilloscope Quick Start Guide...
Page 7: ...1 Introduction Front Panel Features Figure 3 Front Panel Features N1000A Quick Start Guide 7...
Page 8: ...1 Introduction Rear Panel Features Figure 4 Rear Panel Features 8 N1000A Quick Start Guide...
Page 39: ...3 Removable Hard Disk 39 N1000A Quick Start Guide...
Page 47: ...5 Contacting Keysight 47 N1000A Quick Start Guide...