34 Keysight 8478B Thermistor Mount
Maintenance
Performance Tests
4-
SWR Measurement, 15 and 18GHz
Figure 4-2
SWR Measurement, 15 and 18GHz
Procedure:
1. Replace the RF unit, filter and directional detector in the 12GHz test setup
with the instruments shown in
.
2. Repeat steps 1 through 8 to determine the SWR at 15GHz. The SWR at
15GHz should be 1.6:1 maximum.
HP 415E (15GHz)________l.6:1 maximum
3. Move the 8690B START/CW control to 18GHz. Repeat steps 1 through 8 to
determine the SWR at 18GHz. The SWR should be 1.6:1 maximum.
HP 415E (18GHz)________1.6:1 maximum.
SPECIFICATION: SWR at 15 and 18GHz = 1.6:1 maximum
DESCRIPTION:
SWR measurement using standard SWR measurement
techniques.