32 Keysight 8478B Thermistor Mount
Maintenance
Performance Tests
4-
Performance Tests
SWR Measurement, 8GHz
Figure 4-1
SWR Measurement Test, 8GHz
Procedure:
1. Connect the test equipment as shown in
. Make sure all
connections are secured tightly.
2. Set the instrument controls as follows:
SPECIFICATION: SWR at 8GHz = <1.35:1
DESCRIPTION:
SWR measurement using standard SWR measurement
techniques.
8690B/8692A
SWEEP SELECTOR:
CW
FUNCTION:
START/STOP
ALC:
IN
AMPLITUDE MOD:
INT SQ WAVE
START/CW:
8GHz
POWER LEVEL:
Adjust to maximum with the UNLEVELED light