Keysight 42030A Four-Terminal Pair Standard Resistor Set
Operation and Service Manual

2  Calibration and Repair

Introduction

This chapter provides the calibration and repair information for the 42030A 

Four-Terminal Pair Standard Resistor Set.

Calibration

The 42030A is calibrated in the parameters given in 

Table 2-1

 with the 

calibration uncertainties given in 

Table 2-2

 through 

Table 2-7

 and equipped 

with a calibration report, when shipped from the factory.

Table 2-1 

42030A Calibration Parameters

Model Number Nominal Value DC Resistance

Series 

Resistance

and Reactance

Parallel 

Resistance

and 

Susceptance

42031A

1 m

42032A

10 m

42033A

100 m

42034A

42035A

10 

42036A

100 

42037A

1 k

42038A

10 k

42039A

100 k

Summary of Contents for 42030A

Page 1: ...Operation and Service Manual Keysight 42030A Four Terminal Pair Standard Resistor Set...

Page 2: ...e under the same terms by which the software is customarily provided to the public Accordingly Keysight provides the Software to U S gov ernment customers under its stan dard commercial license which...

Page 3: ...for Shipment 7 Keysight Packing Materials 7 Non Keysight Packing Materials 7 Specifications 8 DC Resistance 8 Operating Conditions 9 Dimensions 9 Weight 9 Supplemental Performance Characteristics 10 T...

Page 4: ...4 Keysight 42030A Four Terminal Pair Standard Resistor Set Contents...

Page 5: ...stability resistors and are virtually unaffected by changes in ambient temperature ensuring measurement repeatability and reliability Table 1 1 lists the 42030A contents Table 1 1 42030A Contents Mode...

Page 6: ...een checked for completeness and the instrument has been checked mechanically and electrically The shipping contents should consist of the items shown in Table 1 1 If the shipment is incomplete or if...

Page 7: ...l serial number Non Keysight Packing Materials The following general instructions should be used for re packing with commercially available packing materials 1 If shipping to Keysight Technologies off...

Page 8: ...the factory it meets the specifications listed in this section DC Resistance Table 1 2 lists the nominal DC resistance for each resistance module of the 42030A Table 1 2 DC Resistance Model Number DC...

Page 9: ...onmental conditions listed in Table 1 3 Dimensions Table 1 4 shows the 42030A dimensions Weight Table 1 5 shows the 42030A weight Table 1 3 Operating Conditions Temperature 23 C 5 C Relative Humidity...

Page 10: ...e Conditions The 42030A must be stored or shipped under the ambient environmental conditions listed in Table 1 7 Table 1 6 DC Resistance Temperature Coefficient and Stability Model Number Nominal Valu...

Page 11: ...The 42030A is calibrated in the parameters given in Table 2 1 with the calibration uncertainties given in Table 2 2 through Table 2 7 and equipped with a calibration report when shipped from the facto...

Page 12: ...036A 100 0 02 42037A 1 k 0 02 42038A 10 k 0 02 42039A 100 k 0 02 Table 2 3 42035A 10 Resistor AC Calibration Uncertainty Frequency Calibration Uncertainty Series Resistance Reactance 1 MHz 0 12 0 010...

Page 13: ...0 1 5 S 4 MHz 0 15 2 0 S 5 MHz 0 15 3 0 S 10 MHz 0 4 5 S 13 MHz 0 6 8 S Table 2 6 42038A 10 k Resistor AC Calibration Uncertainty Frequency Calibration Uncertainty Parallel Resistance Susceptance 100...

Page 14: ...s calibration laboratories satisfy all of these calibration uncertainties For complete information on calibration contact your nearest Keysight Technologies sales and service office Calibration Interv...

Page 15: ...This information is subject to change without notice Keysight Technologies 1992 2019 Edition 3 August 28 2019 42030 90001 42030 90001 www keysight com...