*TST?
Performs a basic self-test of the instrument and returns a pass/fail indication. The
self-test is
more comprehensive than the
self-test.
Parameter
Typical Return
(none)
+0 (pass) or +1 (one or more tests failed)
Perform self-test:
*TST?
l
The self-test takes just a few seconds to complete.
l
If one or more tests fail, the query r1 and stores an error in the error queue. For a complete list-
ing of the error messages related to self-test failures, see
l
Following the test, the instrument returns to the instrument state that existed before the self-test.
*WAI
Configures the instrument's output buffer to wait for all pending operations to complete before executing
any additional commands over the interface.
Parameter
Typical Return
(none)
(none)
Wait until all pending operations complete:
*WAI
l
Configuration changes and measurements are considered pending operations. Therefore, *WAI causes
these changes to occur and complete.