Service
1
Keysight 1000B Series Oscilloscopes Service Guide
27
18
Calculate the reference gain as follows:
Record this value in the Calculated Gain @ 1 MHz column of the
“Performance
Test Record”
on page 35.
19
Change the signal generator frequency to the value for the model being tested
as shown in the table below.
20
Change the oscilloscope time base to the value for the model being tests as
shown in the table above.
21
Using the Vpp reading, calculate the Vrms value using the following expression
and record it in the
“Performance Test Record”
on page 35:
For example, if Vpp = 1.24 V:
22
Using the power meter reading, convert this measurement to Volts RMS using
the expression and record it in the
“Performance Test Record”
on page 35:
For example, if Pmeas = 3.65 mW:
Table 7
Oscilloscope Models and Signal Generator Frequency
Setting
Model
DSO1152B
DSO1102B
DSO1072B
DSO1052B
Frequency
150 MHz
100 MHz
70 MHz
50 MHz
Time Base
2 ns/div
5 ns/div
10 ns/div
10 ns/div
Gain
1
MHz
Vout
1
MHz
Vin
1
MHz
----------------------------
=
Vout
max
Vpp
max
2 2
-------------------
=
Vout
max
1.05
2 2
----------
1.05
2.828
-------------
371 mV
=
=
=
Vin
max
P
meas
50
Ω
×
=
Vin
max
3.65 mW
50
Ω
×
427 mV
=
=
Summary of Contents for 1000B Series
Page 1: ...Keysight 1000B Series Oscilloscopes Service Guide ...
Page 6: ...6 Keysight 1000B Series Oscilloscopes Service Guide Contents ...
Page 8: ...8 Keysight 1000B Series Oscilloscopes Service Guide Figures ...
Page 10: ...10 Keysight 1000B Series Oscilloscopes Service Guide Tables ...
Page 40: ...40 Keysight 1000B Series Oscilloscopes Service Guide 1 Service ...
Page 44: ...44 Keysight 1000B Series Oscilloscopes Service Guide Index ...
Page 45: ......