Service
1
Keysight 1000B Series Oscilloscopes Service Guide
23
1
Connect the equipment as shown in
Figure 4
.
2
Preset and calibrate the power meter according to the instructions found in the
power meter manual.
3
Set up the Power Meter to display measurements in units of Watts.
4
Press the
[Default Setup]
front panel key.
5
Press the
[Auto-Scale]
front panel key.
6
Set the channel 1 probe attenuation to 1X.
7
Set the channel 1 vertical scale to 200 mV/div.
Oscilloscope
Power Meter
E4418B
Power sensor
cable
Signal Generator
Power splitter
11667B
Type-N (m) to
SMA (f) adapter
Power sensor
8482A
SMA cable
SMA to BNC adapter
50
W
feed
through
Figure 4
Connecting Equipment for Maximum Frequency Check Test