Installation Note N9040-90027
5
Option FT1 and FT2 - Frequency Mask Trigger Installation Kit
Installation Procedure
License Installation Over USB
1.
Locate the Option Upgrade Entitlement Certificate (5964-5178) from the kit.
2.
Redeem the Option Upgrade Entitlement Certificate by following the instructions on the
Certificate.
3.
After redeeming your Option Upgrade Entitlement Certificate you will receive an email with an
attached License File.
4.
Locate a USB storage device. Perform a virus scan on this device before use.
5.
Save the License File to the root directory of the USB Storage Device.
6.
Connect the USB Storage Device to the signal analyzer USB port. Windows will detect the new
hardware and may display the configuration menu shown in
. This menu may be
configured according to your preferences.
Figure 1
USB Storage Device Configuration Menu