U3020AS62
General Performance
General Performance
The Test Set performance is specified for a standalone instrument.
Power Requirements
•
100/120/220/240 VAC (50/60 Hz)
•
The instruments can operate with mains supply voltage fluctuations up to ± 10% of the
nominal voltage.
•
Air conditioning equipment (or other motor–operated equipment) should not be placed on
the same ac line that powers the Test Set.
•
U3022AS62 maximum power is 50 W.
Environmental Tests
The U3020AS62 complies with all applicable safety and regulatory requirements for the
intended location of use.
•
Operating Environment (Indoor Use)
•
Operating Temperature 0 to 55 °C
•
Storage Temperature: −40 to +70 °C
•
Operating Altitude: 0 to 2000 meters (~ 6,562 feet)
•
The instrument can safely operate in a relative humidity of 80% for
temperatures to 31 degrees C, decreasing linearly to 50% relative humidity
at 40 degrees C.
This product has an auto-ranging line voltage input. Be sure the supply
voltage is within the specified range.
Keysight U3020AS62 User's and Service Guide
7
Definitions
•
Specifications
describe the performance of parameters covered by the product warranty
(temperature 0 to 50 °C, unless otherwise noted.)
•
Typical
describes additional product performance information that is not covered by the
product warranty. It is performance beyond specification that 80% of the units exhibit with a
95% confidence level over the temperature range 20 to 30 °C. Typical performance does not
include measurement uncertainty.
•
Nominal
values indicate expected performance or describe product performance that is useful
in the application of the product, but is not covered by the product warranty.
•
Characteristic Performance
describes performance parameter that the product is expected to
meet before it leaves the factory, but is not verified in the field and is not covered by the
product warranty. A characteristic includes the same guard bands as a specification.